H&L PEEP OFF
The H&L PEEP OFF message is displayed when;
The High Breath Rate alarm (HIGH f) is turned on/set to a value and,
Both the High PEEP and Low PEEP alarms are turned off
This is an informational message only. The message is displayed at power up, when monitored data
is being scrolled automatically, and when no front panel activity is detected for 60 seconds.
To clear the H&L PEEP OFF message:
1) Push any front panel button or turn the Set Value knob.

HIGH f OFF

The HIGH f OFF message is displayed when;
The High Breath Rate alarm is turned off by being set to HIGH f OFF and,
The High PEEP alarm (HIGH PEEP) is turned on.
This is an informational message only. The message is displayed at power up, when monitored data
is being scrolled automatically, and when no front panel activity is detected for 60 seconds.
To clear the HIGH f OFF message:
1) Push any front panel button or turn the Set Value knob.
The HIGH f OFF message is not displayed during SBT mode.

LMV LPPS OFF

The LMV LPPS OFF message is displayed when;
The Low Minute Volume alarm is turned off by being set to dashes and,
The LPP alarm is set to VC/PC ONLY. When this setting is selected, the Low Pressure
alarm applies only to Volume Control and Pressure Control breaths.
This is an informational message only. The message is displayed at power up, when monitored data
is being scrolled automatically, and when no front panel activity is detected for 60 seconds.
To clear the LMV LPPS OFF message:
1) Push any front panel button or turn the Set Value knob.
Page 9-34
®
LTV
1200 Ventilator
Operator's Manual
p/n 18247-001, Rev. B