Page 1
Agilent High Sensitivity D1000 ScreenTape System Quick Guide The Agilent 2200 TapeStation system is an automated platform for simpler, faster and more reliable electrophoresis. It is made up of 3 elements: • 2200 TapeStation system (G2964AA) or 2200 TapeStation Nucleic Acid system (G2965AA) •...
Page 2
Additional Material Required (Not Supplied) • Volumetric pipette • Centrifuge 2200 TapeStation instruments are supplied with an optional IKA MS3 vortexer which N O T E includes a 96-well plate adaptor suitable for both 96-well PCR plates and 8-way strips.
Page 3
C A U T I O N ➔ Use only the recommended consumables and reagents with the 2200 TapeStation system. General Information on Working with DNA • For best results ensure that all reagents are allowed to equilibrate to room temperature N O T E for 30 min prior to use.
Page 4
Parts required Description 5067-5584 High Sensitivity D1000 ScreenTape Launch the 2200 TapeStation Controller Software. Load High Sensitivity D1000 ScreenTape device and loading tips into the 2200 TapeStation instrument. Sample Preparation High Sensitivity D1000 ScreenTape Assay Parts required Description 5067-5585 High Sensitivity D1000 Reagents...