Cleaning and disinfecting
Table 1: Approved cleaners/disinfectants
Product
2
Sani-Cloth AF3
Sani-Cloth Plus
PI-Spray II
1
For maximum effectiveness, the component being cleaned must remain wet with disinfectant for a minimum period
of time.
2
Qualified for use as an intermediate-level disinfectant for mycobacteria.
3
The P11x Transducer is not licensed for use in Canada.
4
The L52x Transducer is for veterinary use only.
Refer to the cleaners and disinfection document available at
current cleaners and disinfectants.
Table 2: High-level compatible disinfectants
Disinfectant
Compatible transducers
Cidex
C8x, C11x, C35x, HFL38xi,
HFL50x, HSL25x, ICTx, L25x,
L38xi, P10x, P11x
Cidex OPA
C8x, C11x, HFL38xi, HFL50x,
HSL25x, ICTx, L25x, L38xi,
L52x
1
The P11x Transducer is not licensed for use in Canada.
2
The L52x Transducer is for veterinary use only.
Refer to the cleaners and disinfection document available at
current cleaners and disinfectants.
Cleaning the ECG cable
This section is no longer applicable and has been removed.
10
Compatible Transducer
C8x, C11x, C35x, HFL38xi,
HFL50x, HSL25x, ICTx, L25x,
L38xi, P10x, P11x
4
HFL50x, L52x
C8x, C11x, HFL38xi, HFL50x,
HSL25x, ICTx, L25x, L38xi, L52x
3
P10x, P11x
, rC60xi, rP19x
1
, rC60xi, rP19x
2
, P10x, rC60xi, rP19x
Minimum wet contact time
3 minutes
3
, rC60xi, rP19x
3 minutes
10 minutes
4
,
www.sonosite.com
Temperature
25°C, 77°F
20°C, 68°F
www.sonosite.com
for a complete list of the most
Disinfectant Soak Duration
45 minutes
12 minutes
for a complete list of the most
Cleaning the ECG cable
1
Need help?
Do you have a question about the SonoSite SII and is the answer not in the manual?
Questions and answers