HBM WE2110 Operating Manual page 67

Digital weighing indicator
Table of Contents

Advertisement

Available languages

Available languages

WE2110
Diagnostic Errors
(E 0001)
The power supply voltage is too low. (check supply)
(E 0002)
The power supply voltage is too high. (check scale / cables)
(E 0004)
The load cell excitation voltage is too low. (check scale)
(E 0008)
The load cell excitation voltage is too high. (check scale)
(E 0010)
The temperature is outside of allowable limits. (check location)
(E 0020)
Count-by setting is too high or low for the set scale capacity
(E 0040)
Excitation sense lines 1-2 are not connected. The error message can be
deactivated with <OPTION><SENS.CH><OFF>.
(E 0080)
Excitation sense lines 3-4 are not connected. The error message can be
deactivated with <OPTION><SENS.CH><OFF>.
Excitation sense lines 1-2 and 3-4 are not connected. The error message can
(E 00C0)
be deactivated with <OPTION><SENS.CH><OFF>.
(E 0100)
The digital setup information has been lost. (restore setup)
(E 0200)
The calibration information has been lost. − (re-calibrate)
(E 0300)
All setup information has been lost − (enter setup and calibrate)
(E 0400)
The HBM factory configuration information has been lost. (service)
(E 0800)
The EEPROM memory storage chip has failed (service)
(E 2000)
Real time clock problem (service)
(E 8000)
The EPROM memory storage chip has failed (service)
The "E" type error messages are additive. For example if a condition is detec-
ted where the power supply voltage is low, resulting in a reduction of excitation
voltage, the resulting Error messages will be E0005 (0001 + 0004).
A2181-1.0 de/en
67
HBM

Hide quick links:

Advertisement

Chapters

Table of Contents
loading
Need help?

Need help?

Do you have a question about the WE2110 and is the answer not in the manual?

Questions and answers

Table of Contents

Save PDF