Electronic error
If the mylife
YpsoPump
detects an
™
®
internal malfunction, an electronic
error is displayed. All the functions of
the pump are cancelled. In the event
of an electronic error the pump dis-
plays two different screens alternately
which show you how to restart the
pump in a few steps:
If the electronic error remains after the pump has been put into operation again or if it reoccurs after a certain
time, stop using the pump, remove the alkaline battery and contact Customer Service.
Disconnect the infusion set from your
body and remove the alkaline battery
from the battery compartment.
Then press the function button for 2
seconds. The pump is now devoid of
a battery and in the storage status.
To put the pump back into operation
perform the steps on page 32 and
check your therapy settings. Change
the cartridge and the infusion set in
accordance with the steps on page
87.