If you need to connect other Thermalert sensors, such as the CI3 and TX, use
the wiring layout in Table B-3.
TERMINAL
NUMBER
6
6
7
8
9
The GP monitor is supplying power to the CI3 and TX when
connected as described in Table B-3, and external power con-
nections are unnecessary. (Maximum power = 24VDC @ 50mA.)
Please note that sensing heads other than those listed above
may need external power if their power consumption needs
exceed what is available from the monitor.
B.3

POWER CONNECTIONS

Please refer to Section 2.3.3 for power connection information.
B.4

OUTPUT CONNECTIONS

Please refer to Section 2.3.4 for output connection information and Section
2.3.5 for information on solid state relay wiring.