Polar Instruments Ltd. Garenne Park St. Sampson TONEOHM 950 Guernsey SERVICE Channel Islands GY2 4AF MANUAL ENGLAND http://www.polar.co.uk Fax: +44 (0)1481 252476 Email: mail@polar.co.uk MAN130–2011...
Page 4
TONEOHM 950 SERVICE MANUAL POLAR INSTRUMENTS LTD. HARDWARE WARRANTY Product Warranty. Product hardware is warranted to be free from defects in material and workmanship during the Warranty Period (as defined below). Product hardware is warranted to conform substantially to Polar’s then current (as of the date of Polar’s product shipment) published user documentation during the Warranty Period.
DECLARATIONS DECLARATIONS European Community Directive Conformance Statement Refer to the Operator Manual to determine conformance with European Community Directives.
TONEOHM 950 SERVICE MANUAL SAFETY WARNING The service instructions contained in this manual are for use by qualified electronic service personnel only. WARNING The LIVE and NEUTRAL lines on this unit are BOTH fused. When the unit is connected to its supply, the opening of covers or removal of panels is likely to expose dangerous voltages.
The unit is designed for use indoors in an electrical workshop environment at a stable work station comprising a bench or similar work surface. Use only the accessories (e.g. test probes and clips) provided by Polar Instruments. The instrument must be maintained and repaired by a skilled electronics technician in accordance with the manufacturer’s instructions.
TONEOHM 950 SERVICE MANUAL SPECIFICATIONS Track Resistance RANGES [Ω], 200mΩ, 2Ω, 200Ω, 20KΩ. Instruments with serial numbers below 005975 have a 2KΩ range instead of 200Ω. ACCURACY ±6% in 200mΩ, ±5% in 2Ω ±4% in 200Ω, ±5% in 20KΩ High sensitivity, uncalibrated.
SPECIFICATIONS Drive Source OUTPUT VOLTAGE 0 to 550mV, adjustable. AC in TRACE, DC in all other ranges. PROTECTION Protected to ± 30V. ENVIRONMENTAL OPERATING CONDITIONS The instrument is designed for indoor use only under the following environmental conditions: Altitude Up to 2000m Temperature +5°...
TONEOHM 950 SERVICE MANUAL SYMBOLS REFER TO MANUAL These sockets are for connecting only Polar Instruments probes and connectors for use as described in the Operator Manual. To prevent damage to this product and to ensure its safe use observe the specifications given in this manual when connecting to terminals marked with this symbol.
Current offset calibration — Version 8 firmware and above............... 2-2 Current Trace ............................. 2-2 Plane Shorts ............................2-3 SECTION 3 – CIRCUIT DESCRIPTION....................3-1 TONEOHM 950 THEORY OF OPERATION..................3-1 Main Operating modes........................3-2 Track resistance ..........................3-2 Track Current ..........................3-2 Trace ..............................
Page 13
TONEOHM 950 SERVICE MANUAL Plane Shorts ........................... 3-2 Detailed Theory of operation......................3-3 Microvolt Amplifier .......................... 3-3 Trace Amplifier ..........................3-3 Current Source ..........................3-3 Plane Amplifier and Vector Discriminator..................3-3 Vectored Plane Stimulus ........................ 3-4 Drive Source........................... 3-5 Tone Generator ..........................
SECTION 1 PERFORMANCE CHECK MAINTENANCE AND CALIBRATION OF INSTRUMENTS For most of Polar Instruments’ products there are two maintenance procedures — the Performance Check and the Calibration. Some instruments may have a single (combined) procedure. Performance Check The Performance Check (or Checkout Procedure) is used to verify the basic functions of the instrument.
TONEOHM 950 SERVICE MANUAL PERFORMANCE CHECK Equipment required • DC voltage source, variable from 1V to 10V and accurate to 0.1%. • High tolerance resistors (0.5%) of the following values: 10KΩ, 1KΩ (serial numbers below 005975 only), 200Ω, 100Ω, 1Ω and 0.1Ω.
Page 16
PERFORMANCE CHECK 20. Slowly turn the DRIVE SOURCE from minimum to maximum (i.e. clockwise). As the control is rotated, check that initially all direction arrow LEDs are on, then the upper LED only, then all LEDs are off. At the same time the tone goes from high to low then off.
SECTION 2 – ADJUSTMENT PROCEDURE ADJUSTMENT PROCEDURE WARNING: Hazardous voltages are exposed on the PCB when the cover is removed. This procedure should only be performed by a technically qualified person aware of the hazards and taking all reasonable care. Equipment required •...
TONEOHM 950 SERVICE MANUAL Current range adjustment (Serial Nº. 012213 and above — below this Serial Nº. there is no adjustment.) 1. Select the 200mA range. Drive source control fully clockwise. 2. Connect the drive source and the DVM across the high tolerance 1Ω resistor.
ADJUSTMENT PROCEDURE Plane Shorts 1. CONNECT the 4 Plane Stimulus leads BLUE to YELLOW and RED to GREEN 2. Push PLANE twice to enter ACTIVE mode. 3. CONNECT the Plane Clip to the Plane Probe. Adjust R85 for a reading of 00.0 ± 3 counts.
SECTION 3 – CIRCUIT DESCRIPTION TONEOHM 950 THEORY OF OPERATION. The TONEOHM 950 consists of the following functional blocks. 1. Microvolt Amplifier Amplifies/attenuates PCB track voltages, and is used in conjunction with the internal current source for the calculation of track resistance and current.
TONEOHM 950 SERVICE MANUAL Main Operating modes. Track resistance Current is injected into the track with the needle probes using the CURRENT SOURCE module, the µV AMPLIFIER conditions the resulting track voltage and displays the result directly on the LCD.
CIRCUIT DESCRIPTION Detailed Theory of operation. Microvolt Amplifier JP11 is the input connection from the probe sockets. Relays K1, 2 and 3 provide ÷ 1, ÷ 10 and ÷ 1000 attenuation setting for the main Amplifier U1. K4 provides a switchable 120Ω input resistance used in 2mV and 20mV.
TONEOHM 950 SERVICE MANUAL The 950 extracts this information as follows: U12 Amplifies the low level signal from the Plane Probe; this chopper-stabilised amplifier is clocked by a signal synchronous with the Vectored Stimulus to minimise clock noise on the a.c. output signal. JP12 connects to the Drive Source/Sensitivity control to allow user adjustment of sensitivity.
CIRCUIT DESCRIPTION In order to protect semiconductor devices, D30 and D34 are Schottky diodes which conduct if the stimulus voltage attempts to rise above approximately 0.4 volts. This will occur if the stimulus is not connected correctly. As D30 or D34 conduct, current no longer flows through L1 or L2, the microprocessor registers NO_PLANE_DS and turns off K14.
TONEOHM 950 SERVICE MANUAL Note that from Serial Nº. 011265 a line filter was fitted and the line voltage selector switch was replaced by links. The power transformer has undergone several changes — check the REPLACEMENT PARTS list before ordering spares.
SECTION 4 – DISMANTLING INSTRUCTIONS WARNING Service of this instrument should only be performed by skilled electronics service personnel. There are hazardous voltages inside the instrument when connected to its power supply. REMOVE THE POWER CORD before touching any part of the line input circuit. Note that high voltages may continue to be present for 2 minutes after power is removed until internal capacitors discharge.
SECTION 5 – LINE VOLTAGE AND FUSE CHANGING WARNING Service of this instrument should only be performed by skilled electronics service personnel. There are hazardous voltages inside the instrument when connected to its power supply. REMOVE THE POWER CORD before touching any part of the line input circuit. Note that high voltages may continue to be present for 2 minutes after power is removed until internal capacitors discharge.
SECTION 6– FAULT DIAGNOSIS WARNING Service of this instrument should only be performed by skilled electronics service personnel. Hazardous voltages are exposed on the PCB when the cover is removed. These procedures should only be performed by a technically qualified person aware of the hazards and taking all reasonable care.
If it becomes necessary to repair the instrument, in the first instance contact the Polar Instruments distributor in your country. In case of difficulty contact Polar Instruments Ltd. at the address at the front of this manual. Do not send the instrument until shipping instructions have been received from the...
Note: Parts marked with an asterisk (*) have been subject to modification in later instruments. Safety critical parts (listed in bold type) must be replaced with parts obtained from Polar Instruments Ltd or your Polar Instruments distributor to ensure continued safe operation.