7#5HPRYDO#)#5HSODFHPHQW#
8. Remove the three hex head (7/64"
allen) screws from the base of the
Tower Assembly and lift the assembly
out.
NOTE: Long screw at base.
Replacement Procedure
Replace by following the above steps in
reverse order.
Verification Procedure
See Verification Matrix,
Section 5.2.
®
CELL-DYN
3200 SYSTEM SERVICE MANUAL
• Tower Assembly (RR-8)
Figure 4-50: Tower Assembly Removal
Part Number 9140300