TSI Model 4040/4043/4045 General Purpose Thermal Mass Flowmeter Operation and Service Manual November 2002 P/N 1980339 Revision E MAIL/SHIP TO: E-mail address: TSI Incorporated Flowmeters tsi.com 500 Cardigan Road Shoreview, MN 55126-3996 Web site: http://www.tsi.com U.S. INTERNATIONAL Sales and Customer Service:...
Page 4
Copyright TSI Incorporated / November 2002 / All rights reserved. Address TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 USA WARNING These flowmeters employ a heated platinum sensor. They should not be used with flammable or explosive gases or mixtures. Caution TSI flowmeters are not medical devices under FDA 510k and in no situation should they be utilized for...
4043 Mass Flowmeter, 0.75 inch inlet/outlet 4045 Computer Cable 1303583 Analog Cable 1303584 Filter 22 mm ISO-taper (for Model 4040) 1602292 3/8” Female NPT (for Model 4043/45) 1602300 AC Adapter 120 V, North America, ungrounded 2613033FLO 8918-NA 100-240 V, NEMA 5-15 plug, grounded...
Page 8
Parts Identification Figure 1-1 Model 4040/4043/4045 Mass Flowmeter 1. On/Off Switch 4. DC Power Input 2. Display 5. Flow Inlet 3. Mounting Inserts (2) 6. Computer Serial Interface and Analog Output and Optional Power Input Connector Chapter 1...
Chapter 2 Setting-Up Supplying Power The flowmeter can be powered in one of two ways: through the power jack using the supplied AC adapter or through the mini-DIN connector. The DC power input connector is shown below along with the power requirements. Power Supply: 7.5 VDC ±...
Connecting Filter and Flow Tubes The Models 4040/4043/4045 have an exposed thermal flow sensor, which must be protected from foreign matter and particles in the gas flow. TSI has supplied a filter, which should be connected to the inlet of the flowmeter, however, any filter will work as long as it has a minimum efficiency of 99.9%.
Configuration Software TSI has several software utilities to help you communicate with your flowmeter to change parameters and to obtain flow data. You can download the latest versions of these at no charge from our web site: http://flowmeters.tsi.com 1. If you only want to change one or more of the operating parameters shown in Table 3 (above), the easiest way is to use the software utility called “TSI setup.”...
Page 12
3. If you plan to develop a more sophisticated program for data collection and control using LabVIEW, you can download a demonstration program called “Real-time Demo Program” and the source code “Source Code for Real-time Demo Program.” This program is intended to be a basic demonstration program and not a practical laboratory tool.
Flowrate Measurement Flowrate data can be obtained from the Models 4040/4043/4045 through the LCD display, RS232 serial port or the linearized analog output. The analog output is a 0 to 10 volt DC linear signal representing 0 to 300 Std L/min (Model 4043: 200 Std.
Pressure Measurement The Models 4040/4043/4045 measure absolute pressure near the outlet of the flowmeter in the SI units of kilo-Pascals (kPa). Pressure measurements are required when converting from standard to volumetric flow. Absolute pressure measurements are displayed on the LCD display and are available through the RS232 serial port.
Remove dust, particles and fibers from the sensor, with clean, dry compressed air. The flow sensor will break if touched. Never run liquids through the flowmeter and never touch the sensor with a brush. Dust or other deposits on the flow sensor will degrade the 4040/4043/4045’s flow accuracy. Caution The flowmeter must be switched off for cleaning.
Chapter 5 Troubleshooting Table 4 lists the symptoms, possible causes, and recommended solutions for common problems encountered with the flowmeter. If your symptom is not listed, or if none of the solutions solves your problem, please contact TSI. Table 4. Troubleshooting Symptom Possible Causes Corrective Action...
Appendix A Specifications Specifications Flow Measurement Measurement Range Models 4040/4045: 0 to 300 Std L/min. Model 4043 only: 0 to 200 Std L/min. Accuracy Air, O : 2% of reading or 0.05 Std L/min, whichever is greater , Air/O mixtures: 3% of reading or 0.1 Std L/min, whichever is greater Accuracy stated at standard conditions: 21.1°C and...
Specifications Physical Dimensions External Dimensions See Diagram Tube Adapters Model 4040: 22 mm male ISO Taper (Inlet & Outlet) Model 4043: ½ inch straight Model 4045: ¾ inch straight Weight 180 grams Flow Body Material PolyCarbonate Computer Interface Connector 8 pin mini-DIN...
Page 21
Pressure Drop - Model 4040 1000 Including external filter No filter 0.01 1000 Flow rate (SLPM) Pressure Drop - Model 4040 Pressure Drop - Model 4043 1000 1000 Including Including external filter external filter No filter No filter 0.01 0.01...
Page 22
Pressure Drop - Model 4045 1000 Including external filter No filter 0.01 1000 Flow rate (SLPM) Appendix A...
P m = Absolute pressure measured in flow tube in units of kPa To demonstrate the difference in output, consider the following. Measured flow displays 100 Std L/min at 15C and 117kPa. The volumetric flow as calculated by the Models 4040/4043/4045 would be as follows. ...