Maintenance
Illustration 33: Removing the cuvette holder.
The cuvette holder can now be cleaned with a mild soap solution.
For disinfection, standard solvent-free preparations can be used.
To replace the cuvette holder, simply push it in the correct
position into the opening in the housing until it engages.
Please wait until the cuvette holder is completely dry before
replacing it.
Do not use any cleaning agent for cleaning the
cuvette holder that could leave scratches on its surface.
WARNING!
6.1.3
Optical unit
The optical unit is situated inside the photometer and has no
direct contact with the cuvette therefore no routine cleaning is
needed. Cleaning the optical unit can become necessary if the
measured maximum light intensity of the photometric light source
no longer achieves the appropriate level required for testing. The
optical unit should be cleaned when the following error message
is displayed:
Dirty Optics – Use Optics Cleaner (see
Illustration 34, Display screen, Measurement Error).
48
HemoPoint
H2 User's Guide
®