Page 1
H A N D B O O K iina70e1-m (1408) Translation of the original handbook Catalog-No. UL 1000: 550 - 000A 550 - 001A 550 - 002A UL 1000 Fab: 550 - 100A 550 - 101A from software version V5.14 UL1000 Fab and UL1000 Helium Leak Detector...
Connections for the Data Acquisition Systems 2.3.2.1 Accessories 2.3.2.2 Digital Out 2.3.2.3 Digital In 2.3.2.4 Recorder 2.3.2.5 RS232 2.3.2.6 Remote Control Vaccum Connections 2.4.1 Inlet Port 2.4.2 Exhaust 2.4.3 Vent 2.4.4 Purge-connection (UL1000 Fab) / Gas ballast (UL1000) 32 Default parameters Table of Contents...
Page 4
Description of the Initial Operation 3.2.1 Start up and Measure 3.2.2 Internal Calibration 3.2.3 Verification Description and Working Principle Introduction Components of the UL1000 and UL1000 Fab 4.2.1 Vacuum System 4.2.2 Control Panel 4.2.2.1 LC Display 4.2.2.2 START Button 4.2.2.3 STOP Button 4.2.2.4...
Page 6
Replacing the Filter Insert 9.10 Monitoring the Oil Level of the D16 B and Topping up the Oil 111 9.11 Oil Change for the D16B 9.12 Scroll Pumps (UL1000 and UL1000 Fab only) Appendix Diagram Index Declaration of Conformity Table of Contents...
This technical handbook contains important informations on the functions, installation, start-up and operation of the UL1000 and UL1000 Fab. General We reserve the right to modify the design and the specified data. The illustrations are not binding.
The range of the preamplifier and the vacuum ranges are selected automatically. The autoranging feature of the UL1000 Fab covers the entire range or leak rates depending on the selected operating mode. Not only the leak rate signal, but also the pressure in the test sample (inlet pressure P1) and the forevacuum pressure (P2) are used for control purposes.
Page 9
This function determines the internal ZERO level of the unit which is then deducted from the currently measured leak rate signal. This function is enabled by pressing the Start key provided the UL1000 or UL1000 Fab have run at least 20 s in "Standby" mode or in "Ventilation" mode.
Support from INFICON Service If an instrument is returned to INFICON or an authorised representative of INFICON, please indicate wether the instrument is free of substances damaging to healths or wether it is contaminated. If it is contaminated also indicate the nature of the hazard.
1.2.1 Service Centers In case you urgently need assistance please get in touch with the local INFICON Service in your country or the service hotline in Cologne, Germany: Algeria jhj@agramkow.dk Finland jhj@agramkow.dk Agramkow Phone: +45 741 236 36 Agramkow Phone: +45 741 236 36...
Great caution is necessary when using hydrogen! No smoking, no naked flames, avoid sparks. Caution The UL1000/UL1000 Fab is only allowed to be used as a leak detector. It may not be used as a pump system (especially not for pumping off aggressive or humid gasses)
Page 14
If you plan to detect noxious matters please contact the manufacturer. Rules for decontamination will be developed then. If the leak detector alreay has been in contact with dangerous gases please fill the declaration of contamination, too, and send it to INFICON before you send the parts. General Information...
Run-up time (after starting) Hinweis To get down to the minimum detected leak rate range some conditions must be fulfilled: • UL1000 and UL1000 Fab has fully warmed up • Ambient conditions must be stable (temperature, no vibration/accelerations.) • The part under test has been evacuated long enough (background is no longer decreasing) •...
80% non condensing Max. permissible height above sea level (during operation) 2000 m Unpacking Unpack the UL1000 and UL1000 Fab immediately after delivery, even if it will be installed later on. Examine the shipping container for any external damage. Completely remove the packaging materials.
– Spare Parts List UL1000 and UL1000 Fab • hooks to wrap power cord (with screws) (see arrow 6) • Tool to open the UL1000 and UL1000 Fab (see arrow 7) • O-Ring with filter (for use at applications with particles) Fig. 1-2...
LeakWare (software package) 14090 1.4.2.1 Sniffer line SL200 By use of the sniffer line the UL1000 and UL1000 Fab can easily be converted to a sniffer leak detector. The length of the sniffer line is 4m (i.e. 12 feet). 1.4.2.2 Toolbox The toolbox is a detachable compartment with a lockable lid.
The RC1000 is a wireless remote control that allows to operate the UL1000 and UL1000 Fab from distance up to 100 m. It provides the functions START, STOP/ VENT, ZERO and speaker volume, and displays leak rate in bargaraph or in chart mode.
1.4.2.6 Test chamber TC1000 This test chamber turns the UL1000 / UL1000 Fab into a workstation to test hermetically sealed components. Testing according to MIL-STD 883 can be done easily, fast and accurate. The test starts automatically when the chamber lid is closed, test parameters like cycle time and rejectant level can be setted in the menu Auto Leak Test (see 6.6.1.6).
Caution When transporting over longer distances the original packaging must be used. The castors must not be fixed when the UL1000 and UL1000 Fab is shipped in a crate. UL1000 Fab with Triscroll TS 620 For transportation the chassis plate where the pump is mounted on has to be secured by a transportation fixing.
Page 22
Fig. 2-1 For transportation fixing the screws are tightened to the chassis plate. For operation of the UL1000 Fab the screws should be loosened. To loosen the screws first loosen the counter nut that is accessible from underneath: Fig. 2-2...
Fig. 2-3 For transportation tighten the screws again and fix them by the counter nuts. Working Location Move the UL1000 and UL1000 Fab to the desired position and arrest the castors. Danger S TOP Caution: Exhaust gases and fumes: Exhaust gases and fumes from oil-sealed pumps may be harmful to health.
Page 24
It is recommended that you check all major helium sources in the vicinity of the UL1000 and UL1000 Fab within a radius of about 10 m for the presence of any big leaks. You may use the sniffer probe for this.
(Circuit breaker 16A max. according to IEC/EN 60898 with tripping characteristic B). The mains voltage rating for the UL1000 and UL1000 Fab can be read off from the name plate beneath the mains socket Fig.
Page 26
Fig. 2-4 secure fixture power cord Notice: If the machine is not operating the cable can be stored at the cable holders. Fig. 2-5 storing power cord Installation...
7. Mains Socket 8. Mains Switch 9. Hole to mount cable hooks. 10. Speaker 11. Vent 12. Purge (UL1000 Fab) / Gas ballast (UL1000) Fig. 2-6 Tipp The sockets: Accessories, Digital Out, Digital In and Recorder have pin 1 on top. The pin numbers are counted downwards. The socket 2 and 3 are coded mechanically to avoid a confusion with the counter plug.
2.3.2.1 Accessories The following accessories may be connected to the sniffer line SL200 (Please refer to Chapter Fig. 2-6/1) or the test chamber TC1000: Contact pins 1 and 3 are fused with a 0.8 A slow-blow fuse. The amount of power which can be drawn is limited to 10 W.
Closed in case recorder output is invalid. Only used when record output is set on „leak rate“. 2.3.2.3 Digital In These inputs can be used to control the UL1000 and UL1000 Fab with a programmable logic control (PLC). Assignment +24V, bridged with pin 1 of socket „Digital Out“...
UL1000 and UL1000 Fab and the chart recorder from the same mains phase. If this is not possible, you must make sure that the frame ground of both instruments is kept at the same potential.
Fig. 2-6/6 is a serial interface to control the UL1000 and UL1000 Fab by the RC1000. The RC1000 Remote Control can be connected via the wireless transmitter or via an extension cable with a RJ45 plug. Refer to the RC1000 Technical Handbook for more information.
Vaccum Connections 2.4.1 Inlet Port The inlet port is located on the top of the UL1000 and UL1000 Fab. The size of the flange is DN 25 KF. Warning Risk of injury due to sucking connection flange. If the Vacuum-Mode of the UL1000 is activated, the connection flange may suck bodily parts around the connection flange.
Default parameters The following parameters are set like shown when in the menu of the UL1000 and UL1000 Fab under Settings Parameters load/save, „load default values“ is chosen. Auto-scaling: Scaling logarithmic Display range: 4 decades Time axis: 32 seconds...
First Operation Check The steps for an initial operation are described in this chapter. It is explained how to switch on the UL1000 and UL1000 Fab, how to measure and how to carry out an internal calibration. Notice: If anything unexpected happens during the initial operation or the leak detector acts in a strange way the UL1000 and UL1000 Fab can be switched off by the mains switch at any time.
Page 35
The start up procedure takes about 3 minutes and the end is indicated by a beep. The UL1000 and UL1000 Fab is in Stand-by mode now. Fig. 3-1: Top view of the Standards Pos.
3.2.2 Internal Calibration Proceed the internal calibration (Please refer to Chapter 7.2.1 Internal Calibration). For better quantitative measurements please allow the unit to warm up (15 … 20 minutes). • Press Calibration (Soft Key no. 5 Fig. 3-2/8) to get into the calibration menu. •...
Fab. In addition the vacuum can be generated by pumps with are set up in parallel to the leak detector. Another operating mode of the UL1000 and UL1000 Fab is the Sniff mode which can only be used when a sniffer line (See Chapter 1.4.2 Accessories and...
4.2.1 Vacuum System The vacuum diagram below shows the major components inside the UL1000 and UL1000 Fab: Fig. 4-1: Vacuum Diagramm UL1000 and UL1000 Fab Pos. Description MS: Mass Spectrometer, Helium sensor (180° magnetic field mass spectrometer) Turbomolecular Pump (TMP, provides high vacumm conditions in the MS) V1a …...
Besides maintaining the pressure in the mass spectrometer the pump system is used to evacuate the test parts. It is made sure that the pressure in the mass spectrometer is low enough under all circumstances. The valves V1a, V1b, V2, V4a, V4b control the gas flows when measuring.
START Button Pushing the START Button Fig. 4-2/6 enables the UL1000 and UL1000 Fab to start the measure procedure. If the START button is pushed again in measurement mode, the maximum leak rate indicator („hold“ function) is acitvated. This indicator shows the maximum leak rate since „START“.
Page 42
Fig. 4-3 decreasing background When the measurement signal declines below the saved underground value the underground value will automatically be equated with the measurement signal. As soon as the measurement signal is increasing again the saved decreasing value remains constant. Increasings of the signal are displayed clearly as a leak. Fig.
4.2.2.5 MENU Button When pressing the MENU button (Fig. 4-2/12) the selecting menu is shown at the display. This function is not depending on the operating mode when calibrating. If the menu is opened during the current session the operator will lead to the last screen before the menu was left.
Page 44
Example To change the trigger level 1.0x10 mbar l/s to 3x 10 mbar l/s please press 2/3 (Soft Key no. 3) Fig. 4-5. Fig. 4-5: Numerical entry of the Trigger Level 1 In the submenu press 3 (Soft Key no. 4) Fig.
START Button Valves V1a and V1b open and the sample is pumped down by the roughing pump (UL1000) or scroll pump (UL1000 Fab). At the same time valve V2 is closed to avoid an unacceptable pressure increase in the turbo pump and the mass spectrometer.
Page 46
Since the scroll pump continues to evacuate the test sample the inlet pressure p will continue to drop. Below 2 mbar the UL1000 and UL1000 Fab will switch to FINE mode, i.e. V4a will open and V1b will close. The gas stream enters the turbo pump at an intermediate level.
Fast test results within seconds are achieved by using the internal test leak of the UL1000 or UL1000 Fab for a dynamic calibration, matched to the required test cycle. Leak rates in the 10-9 mbar l/s range can be detected within 5 seconds.
The displayed leak rate corresponds to the helium background concentration in the part under test. Since the UL1000 and UL1000 Fab continues to pump down the part this background leak rate will further reduce. As soon as the leak rate is low enough in respect to your requirements you may start spraying Helium to search for possible leaks.
(see Chapter 5.3.1) 5.3.1 Purging Every time when the UL1000 Fab changes into stand by mode it can start purging automatically after 20 seconds. During this purging the scroll pump is flushed through purge connection (See Fig.
(Please refer to Chapter Mode). ULTRA • Vacuum area Depending on the inlet pressure the UL1000 and UL1000 Fab may be in GROSS, FINE or ULTRA, which is indicated here (Chapter 4.3 Working Modes) ZERO •...
(Please refer to Chapter 6.2.3 Time axis). The intensity axis (y-axis) is defined the same way as the bargraph (Please refer to Chapter 6.2.1 Scale linear/ logarithmic ff). Fig. 5-2: Display, trend mode Operation of the UL1000 and UL1000 Fab...
Mode Please refer to Chapter 6.3. Settings of units, trigger levels and alarms. Trigger & Alarms Please refer to Chapter 6.4. Calibration of the UL1000 and UL1000 Fab. Calibration Please refer to Chapter 6.5. Settings of internal machine parameters. Settings Please refer to Chapter 6.6.
Page 53
Alarm delay Audio alarm type manual internal automatic Calibration (See 6.5 ) Edit leakrate external Start Automatic purge (UL1000 Fab only) Vent delay Vacuum settings Vacuum ranges Leak rate internal test leak Machine factor Background Suppression Zero & Background Zero Mass...
View • Main Menu > View In this menu Fig. 6-3 all features that influence the way data are displayed are put together. Fig. 6-3: The View Menu Key No. Name Description Back Return to the main menu. Settings for bargraph and trend mode. Scale linear/logarithmic Please refer to Chapter 6.2.1.
6.2.1 Scale linear/logarithmic • Main Menu > View > Scale linear/logarithmic These settings apply to the bargraph (= bar underneath the digital figures in the measurement mode) and Y-axis in the trend mode. The scale of the bargraph can either be linear or logarithmic. With the arrows (up and down) it can be determined how many decades the bargraph covers.
Switch off the UL1000 and UL1000 Fab and turn it on again. During the run- up phase press the key no. 3 or 7 so long until the display can be read properly again.
6.2.5 Background in Stand-by • Main Menu > View > Background in Stand-by The internal background leak rate can be displayed in Stand-by mode (ON) or not (OFF).The default setting is OFF. Softkey 3: Off The background leak rate will not be shown. Softkey 5: ? Help Softkey 7: ON...
6.2.7 Lower display limit • Main Menu > View > Lower display limit This parameter defines the lower leak rate limit in the measurement ranges.This is valid for vacuum modes only. Softkey 3, 7: Changing of the lower detection limit between 1×10 and 1×10 Softkey 5: Help...
6.3.1 Auto Leak Test • Menu > Mode > Auto leak test After selecting the Auto leak test mode this message pops up: Fig. 6-5 By pressing OK button the settings menu is called (see 6.6.1.6 ) Notice: If the UL1000 requests a calibration due to the mode change a calibration message will come up.
Page 60
Shut down When the test chamber should remain under vacuum after switching off the UL1000 or UL1000 Fab, push the button (cup of coffee), follow the instruction on the display and switch off the leak detector. This function can also be used when a test of parts should be interrupted due to a brake.
Trigger 1 and Trigger 2 are programmable switching thresholds. When these thresholds will be exceeded the UL1000 and UL1000 Fab reacts as follows: Display In the status line of the display the signs for Trigger 1 and Trigger 2 are displayed inverted if the leak rate exceeds (becomes higher than) the programmed value.
6.4.2 Trigger Level 2 • Main Menu > Trigger & Alarms > Trigger Level 2 The value of the second trigger level can be set. Please refer to Chapter 4.2.2.7 Numerical Entries for the description of the entry. If Trigger 2 is exceeded the corresponding relay will switch. This is also indicated at the display (see above).
6.4.4 Units • Main Menu > Trigger & Alarms > Units The prefered leak rate unit can be selected. There is the choice of 4 (mbar, Pa, Torr, atm)pressure units and 5 leak rate units (mbar l/s, Pa m /s-1, Torr l/s, atm cc/s). Notice: In Sniff mode the following measuring units are electable (Refer to chapter 6.3): ppm, g/a eq (helium leak rate is equivalent with leak rate R134a), oz/...
6.4.6 Audio alarm type • Main Menu > Trigger & Alarms > Audio alarm type The trigger of the audio alarm can be switched on or off. Softkey 2: Pinpoint 6.4.6.1 Use this function to localize a leak with a well-known size. Please refer to Chapter Softkey 3: Leak rate prop.
6.4.6.3 Setpoint The tone is off as long as the leak rate is below the Trigger level 1. Above Trigger 1 the tone varies proportional to the leak rate Fig. 6-11. 3000 Hz - 375 Hz - no tone 1000 • TR.1 Tr.
(local, RS232, PLC). See Chapter 6.6.4 Change rarely necessary settings (Date, Miscellaneous language). See Chapter 6.6.5 Parameter save / load See Chapter 6.6.6 Choose functions of protection of the UL1000 and Monitoring functions UL1000 Fab. See Chapter 6.6.7 Description of the Menu...
UL1000 Fab will just change to Stand-by mode. When the STOP button is pressed for a period of time which is longer than the delay time specified here, the UL1000 and UL1000 Fab will vent the inlet port. Description of the Menu...
Softkey No. 3: FINE only In this mode the UL1000 and UL1000 Fab remains after falling below 2 mbar at the inlet flange in FINE mode. Valve V1a will be closed. When the pressure at the inlet flange is increasing > 1 mbar the UL1000 and UL1000 Fab switches immediately into evacuation mode.
The machine factor can be estimated by taking into consideration the Helium absorbing capability of the UL1000 and UL1000 Fab and the external pump. Exactly, this is the result of the measured leak rate of an external test leak on the test sample once with and then without the external pump.
Page 70
PASS even if the maximum value of the measurement period is not reached. Fig. 6-14 Measurement period Examples for time settings (using the Inficon Test Chamber with a volume of 430 cm³): Range of rejectant level Measurement Period 10E-5...
Page 71
Trigger level 1 The rejectant level for a part to be tested can be set in the range from 10 mbar l/s. Fig. 6-15 Series error messages The number of failed parts in series can be set from 1 to 9. In disable mode this function is switched off.
Page 72
Part under test The number of the first part to be tested can be entered. The number counts automatically up at the next test cycle. In disable mode this function is switched off. Fig. 6-17 Reference measurement Fig. 6-18 This mode can be used to clean up the chamber after a helium contamination or after a series of failed parts.
> Background Suppression By this mode the internal helium background of the UL1000 and UL1000 Fab will be subtracted at every measurement after pressing START. This function helps to save clean up of the UL1000 and UL1000 Fab after a helium contamination.
Softkey 2: Zero at ULTRA With “Zero at ULTRA” the ZERO functions executes automatically as soon as the measuring range ULTRA is reched for the first time after START. In this mode the ZERO function also can be executed manually via the ZERO button. Softkey 3: Disable: ZERO function Softkey 5:...
Interfaces > Control Location Softkey 2: The UL1000 and UL1000 Fab is controlled via the Digital In connector (See Chapter 2.3.2.3). The START, STOP and ZERO buttons at the control panel are locked. Softkey 3: RS232 The UL1000 and UL1000 Fab is controlled via RS232 interface by an external computer.
Please execute the function „STORE DATE“ in the operating mode „Single Part Measurement“ for starting the record of the measured values. Softkey 7: ASCII Gives the chance to use the UL1000 and UL1000 Fab via a RS232 terminal. Description of the Menu...
6.6.4.3 Recorder output • Main Menu > Settings > Interfaces > Recorder output The signals to be recorded can be selected in this submenu. Softkey 1: Cancel Return to the previous menu without any changes of the present settings. Softkey 2: Arrow up Adress recorder output 1 or 2 Softkey 3: Arrow down Adress recorder output 1 or 2...
LR lin The leak rate output voltage is scaled linear. The fundamental voltage is 0-10 V in scalable steps from 0.5 to 10 volts per decade. For information about scaling see chapter 6.6.4.4 LR log The leakrate is recorded on a logarithmic scale. The voltage output ranges from 1 …...
6.6.5 Miscellaneous • Main Menu > Settings > Miscellaneous The actual date and time, the prefered language and the mains frequency can be set in this submenu. Softkey 2: Time&Date 6.6.5.1 Please refer to Chapter Softkey 3: Language 6.6.5.2 Please refer to Chapter Softkey 4: Leak rate filter 6.6.5.3...
50 Hz for 230 V and 60 Hz for 115 V. Softkey 3: 50 Hz The UL1000 and UL1000 Fab will be run at a mains frequency of 50 Hz. Softkey 6: 60 Hz The UL1000 and UL1000 Fab will be run at a mains frequency of 60 Hz.
Service message exhaust filter The exhaust filter must be maintained at regular intervals to ensure the correct function of the UL1000 and UL1000 Fab. If the service message is activated, the UL1000 and UL1000 Fab reminds you of the required maintenance.
If the calibration request is switched on, a corresponding message will appear when 30 minutes have elapsed after power on or if the temperature of the UL1000 and UL1000 Fab has changed by more than 5 °C (9 °F) since the last calibration. Particle protection •...
Page 83
Monitoring functions > Contamination protection If this mode is switched on the UL1000 and UL1000 Fab closes all inlet valves as soon as the measured leak rate exceeds the programmed leak rate. Thus no more Helium gets into the mass spectrometer. Helium that has gotten into the tool under test can be pumped away by the tool pump.
Page 84
Softkey 3 and 6: Setting of the maximal pressure: upper limit 2 mbar Softkey 4 and 7: Setting of the minimal pressure: lower limit 0.02 mba Softkey 5: Help Maximum evacuation time • Main Menu > Settings > Monitoring functions >...
Information on measured internal data is provided on 4 screens. Softkey 4: Vacuum diagram The vacuum diagram of the UL1000 and UL1000 Fab is shown. Here you can see which valves are opend or closed momentarily and more. Softkey 5: View error list The list of occured errors and warnings will be displayed.
Access Control • Main Menu > Access Control With this menu you can deny or allow access to specific functions of the UL1000 and UL1000 Fab. Fig. 6-23: The Access Control Menu Softkey 2: Access to CAL function 6.8.1 Please refer to Chapter Softkey 3 Access to Trigger&Alarme menu...
> Change Device PIN The access to the UL1000 and UL1000 Fab can be restricted by a Device-PIN. If the Device-PIN is not 0000 the UL1000 and UL1000 Fab will ask for this PIN directly after power on. Without device-PIN the UL1000 and UL1000 Fab foes not even switch on the pumps.
Although the UL1000 and UL1000 Fab is a very stable instrument a calibration is recommended from time to time to make sure that ambient temperature changes or dirt or other impacts don’t adulterate the measurements.
7.2.1.2 Manual Internal Calibration When Manual Internal Calibration is selected it is assumed that the UL1000 and UL1000 Fab is connected to a component under test (if not please go to Automatic Internal Calibration). After starting the Manual Internal Calibration...
Page 90
• Make sure that the test leak is connected and opened. • Check the leak rate printed on the test leak and compare it with the leak rate at the display. If the leak rates are not identical press Edit leak rate (Soft Key no. 4) and correct the value.
Page 91
• No action required. Fig. 7-4: External Calibration, Step 4 • No action required. Fig. 7-5: External Calibration, Step 5 • Close the external leak standard and confirm with OK (Soft Key no. 8). Fig. 7-6: External Calibration, Step 6 Calibration...
Page 92
If so please press OK (Soft Key no. 8). Fig. 7-7: External Calibration, Step 7 • The UL1000 and UL1000 Fab displays the old and the calculated new calibration factor. Fig. 7-8: External Calibration, Step 8...
Factor of Calibration - Range of Values To avoid a faulty calibration the factor of calibration is tested for plausibility at the end of the calibration routine: When the new factor of calibration is not considerably higher or lower (< factor 2) than the previous factor of calibration the new factor will be accepted automatically.
(e.g. a request for calibration or a service timer has expired). The UL1000 and UL1000 Fab will indicate a message on the LC display and will remain in the Stand-by or the measurement mode.
List of Errors & Warnings The following pages contain a list of all errors and warnings displayed at the control panel. Warning messages are indicated by numbers with a leading W. Error messages are indicated by numbers with a leading E Displayed Message Description and possible solutions E04 Temperature...
Page 96
W46 Emission for cathode 2 can Signal MSIBE on MSV board is not active. Emission for cathode 2 can not can not be switched. be switched on. UL1000 and UL1000 Fab switches to cathode 1. Order a new ion source. Error And Warning Messages...
Page 97
The maximum limit can be set by the menu. Default value is 1.0 mbar. 6.6.1.3. E73 Emission off (p2 too high) PV >> 0.2 or 3 mbar due to an inrush, e. g. The UL1000 and UL1000 Fab will again try to resume the measurement mode. W76 Maximum of evacuation time •...
Page 98
Displayed Message Description and possible solutions W77 Peak not in Range The signal maximum has shifted to mass range alignment limits. • Signal of leak rate was instable during mass adjustment. Calibrate again. • Check the basic setting for the anode voltage through the service menu.
As soon as the different maintenance intervals are reached, a warning is displayed on the screen of the leak detector of the UL1000 and UL1000 Fab/UL1000 each time it is switched on. This message will remain on the display until the maintenance interval has been acknowledged.
The pump has been specified for use in the leak detector UL1000 with Arctic Oil and must for this reason be filled with Arctic Oil (Part No. 20099091) only. When using any other kind of oil, INFICON GmbH Cologne must reject all warranty claims relating to the backing pump.
09.fm (1408) technical handbook Maintenance Plan Subassembly Maintenance work Operating hours/years Service Spare part level UL1000 / UL1000 Fab 1500 4000 8000 16000 24000 number Vacuum system Backing pump D16 B Check oil level, if required exchange I & II...
The 4000 hours maintenance work should be done by an INFICON service technician or a person authorised by INFICON, were by this work should be done once a year. The operating agent reservoir of the turbomolecular pump and the oil of the backing pump in the UL1000 must be exchanged at least once a year SplitFlow 80.
For the scroll module of the Agilent scroll pump, the „tip seal“ must be replaced by an INFICON service technician as soon as 8000 operating hours are attained. If the „tip seal“ is not replaced, it will become necessary to replace the scroll module after 12000 operating hours.
In this case the turbopump SplitFlow 80 will be replaced by a refurbished pump. Description of the Maintenance Work Modifications to the UL1000 and UL1000 Fab/UL1000 which exceed the scope of normal maintenance work must only be performed by trained personnel.
Side wall separator (Plastic wedge) from the accessories. Danger S TOP Before removing one of the covers of the UL1000 and UL1000 Fab/UL1000 the leak detector must be disconnected from the mains power! • Vacuum components which are fitted to the inlet of the UL1000 and UL1000 Fab/ UL1000 must be disconnected from the inlet system.
Replacement filter insert P/N 200 000 685 Danger S TOP Before removing one of the covers of the UL1000 and UL1000 Fab/UL1000 the leak detector must be disconnected from the mains power! • To open the leak detector, refer to Chapter 1.6.1...
Page 107
Fig. 9-2 Replacement of the air filter in the chassis wall Item Description Item Description Air filter cut out Air filter guide Air filter insert Ejection bore (rear) • When inserting the new air filter insert, note the direction of the air flow. It is indicated in the figure above Fig.
Replacing the Exhaust Silencer Required material Replacement silencer P/N 200 99 183 • Switch the UL1000 Fab leak detector off. • Unscrew the silencer from the connection adapter and replace it by the new silencer and tighten it. For this refer to Fig.
1.6.1. Danger S TOP Before removing one of the covers of the UL1000 and UL1000 Fab/UL1000 the leak detector must be disconnected from the mains power! • Loosen the hexagon screw at the bottom of the reservoir vessel and drain the oil into a suitable vessel.
Page 110
Fig. 9-4 Installation location of the exhaust filter Item Description Item Description Clamping ring KF 16 Maximum oil level mark Filter insert Oil level viewing glass Oil reservoir vessel Maximum oil level mark To exchange the filter insert proceed as follows: •...
To open the instrument, observe the instructions given in Chapter 1.6.1. Danger S TOP Before removing one of the covers of the UL1000 and UL1000 Fab/UL1000 the leak detector must be disconnected from the mains power! Maintenance Work...
Required tools Side wall separator Through the oil level viewing glass of the backing pump D16B the oil level and the colour of the oil can be checked visually. The oil level of the vacuum pump must be within the minimum and maximum marks. For this refer to Fig.
Page 113
Further oil changes should be done before and after storing the pump for a longer period of time. Caution Exchange the oil always while the vacuum pump is still warm and with the vacuum pump switched off! Required tools Side wall separator Hexagon socket screw key, size 5 mm;...
Scroll Pumps (UL1000 and UL1000 Fab only) The maintenance intervals of the scroll pump (Agilent TS620) is given in the maintenance plan detailed in 9.4. We recommend to have this maintenance work done by the INFICON Service or a service partner authorised by INFICON. Maintenance Work...
Access Control Hand Set Accessories – Helium leak rate alarm Humidity Ambient temperature Audio alarm I CAL Automatic Internal Calibration INFICON Service information background Inlet area background suppression built-in leak standard Inlet Port Inlet port Inlet pressure Calibration Installation calibration...
Page 117
Power Trigger • consumption Trigger alarm pressure turbo pump Protection turbomolecular pump • Type of Type of protection Pumping speed Purge Purging ULTRA Purpose ULTRA only QT100 Vacuum Diagramm Vacuum mode Valves Recorder Vent Relay Output volume RS232 run-up mode Run-up time Weight Safety Symbols...
Declaration of Conformity Declaration of Conformity...
Page 119
INFICON GmbH, Bonner Strasse 498, D-50968 Cologne, Germany UNITED STATES TAIWAN JAPAN KOREA SINGAPORE GERMANY FRANCE UNITED KINGDOM HONG KONG Vi s i t o ur w eb s i t e f o r c on ta c t i nf o rma t i on an d o t he r s al es of f i c e s w or l dwi de . w w w. i n fi c o n . com...
Need help?
Do you have a question about the UL1000 Fab and is the answer not in the manual?
Questions and answers