Notes for use in legal metrology. For instruments the conformity assessment procedure has been performed (verified balances) Note regarding device operation When individual buttons are displayed, they should be pressed. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
1.3 Application Advice/Technical Support Contact addresses for application advice and our technical support can be found online at: http://www.sartorius.com Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Do not operate the balance if its housing or AC adaptor is damaged. Immediately disconnect the damaged device from the power by pulling the plug. Do not expose the balance, its AC adaptor or the accessories supplied by Sartorius to extreme temperatures, aggressive chemical vapors, moisture, shocks, vibrations or strong electromagnetic fields.
It was developed specifically for the exact determination of the mass of materials in liquid, paste, powder or solid form. Appropriate containers must be used for each type of sample material. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Grip, side panel (draft shield) USB port for mini AB ® Level indicator (Quintix model) DC jack Touchscreen Fastening point for "Kensington" key lock (optional) Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Weighing pan: d 50 mm Shield disk (normal / high) AC adapter with country-specific power plug In-use dust cover Dust cover Installation and operating instructions Application guide * optional Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Insert the panel on the top into the guide rails of the draft shield. Press the panel gently upwards, while simultaneously pressing downwards into the slot of the draft shield. The panel must click into place in the slot. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 11
Insert the weighing pan into the opening of the collection container. Position one of the two shield disks on the collection container. Push the upper and side panels forward to close them. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
You should therefore let a device that has °C been disconnected from its power source acclimatize for approximately 2 hours before reconnecting it to the power. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Hold the balance under the housing, lift it up carefully, and carry it to its new location. The balance needs to be re-leveled and then adjusted each time its setup location is changed. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Push (1) and slide (2) the power plug adaptor (A) required for your power supply into the opening of the AC adaptor module (B). When doing this, the AC adaptor needs to lock into position. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Unlock (1) the AC adaptor and then detach (2) it. Power Connection/Safety Precautions – Only use original Sartorius AC adaptors. The AC adaptor has an IP rating of IP40 in accordance with EN 60529. – Make sure that the voltage rating printed on the manufacturer's ID label is identical to that of your local mains voltage.
The date is displayed in the order of year, month and then day. (The time is displayed in 24-hour format with this setting.) Touch to confirm. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 17
Follow the instructions in Chapter “Leveling the Balance” on page 20. If required, you can subsequently modify your selection in the “Settings” menu (see Chapter “Accessing the Menu and Changing Settings” on page 27). Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
A “Kensington” key lock can be installed at the fastening point on the back of the balance if required. The balance can also be secured at the installation location with a chain and lock, for example. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
To switch the balance on again: Select on the display. The balance starts in the application most recently used before the it was last switched off. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
When the level indicator animation turns green, the balance is correctly leveled. Leveling To exit the function, select The balance has been leveled. Complete the leveling by calibrating the balance. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
If the air bubble is too far to the left, twist the left leveling foot counterclockwise and the right leveling foot clockwise. Check the level indicator after each adjustment and correct if necessary. Complete the leveling by calibrating the balance. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
13 Field for status and warning displays, and activation of isoCAL and leveling Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Back: Run the most recent application. Setup menu: Access the balance system settings (see “System Settings (Setup Menu)” on page 27). Access the balance leveling function via electronic level indicator (Secura ® only – see “Leveling with Electronic Level Indicator (Secura®)” on page 20).
Use the ABC key to switch the keyboard display from numbers to letters. Use the backspace key to delete the last character. key ends the process and saves the characters entered. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
The balance is automatically tared after each component is weighed. The weight value of an individual component or the total weight can be displayed as desired. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 26
The weight of a counted reference sample is calculated and then the objects with an unknown piece count are weighed. The balance displays the number of parts and the piece weight. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Select to confirm your changes. Select to return to the menu. The changed settings are active when you reach the first level of the menu again. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
(Setup) key in the menu. y The Settings window appears. Settings Language Language Select or the top entry in the list. Date and time Device information Calibration/Adjustment Weighing Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
ISO format.) 12h (AM/PM) – The time is displayed in 12-hour format. The hours before noon are displayed with and the hours after noon are displayed with Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
You can configure how the automatic calibration/adjustment function isoCAL starts. Calibration/Adjustment You can view the saved reports on recently conducted adjustments, including the isoCAL deviations detected, and send the output to a PC or a Sartorius laboratory printer. Calibration report isoCAL Under Select the desired start option for isoCAL.
Page 31
Balance not leveled –> Red marking on “LEVEL” display element – Balance not adjusted due to temperature or time –> Red marking on “isoCAL” display element Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 32
Zero/Tare at power on – : If this option is enabled, the balance is automatically zeroed or tared when switched on. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Value without ID : Only if being transmitted to a PC (menu items: USB port: Device/protocol: PC – xxx) Value or date, value Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
: The sample ID is deactivated. To enter and delete numbers and text, e.g., for the sample ID, see “Entering Figures on the Number Pad” on page 24. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
: The balance transmits the data via keyboard command (keyboard emulation) to the currently opened application on the PC in text format. – : Data transmission is deactivated. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 36
Requirement: For PC applications, such as MS Excel, the keyboard must be set to English (USA). You can find more information in Chapter “USB Port” on page 78. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Visit www.scaleman.com your Authorized Sartorius Distributor System Settings (Setup Menu) 2) Connection to Sartorius Lab Printer via the Supplied Printer Cable Device/Protocol Under Select the desired setting for the connection to the printer. YDP30/YDP40 – : The connected printer is detected automatically and the connection is established.
Under Use this function to register the balance memory as a USB removable data carrier on a PC. This function is required by the Sartorius Service Center or when installing the PC driver for the virtual COM port. Further information can be found in Chapter “USB Port” on page 78.
Page 39
Accessing the menu or unlocking it again: Hold down the Menu key for longer than 10 seconds. Supervisor lock Unlock menu? Then select the required function. Menu Unlock Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Until stability is reached, the measured value is shown in gray on 14.42000 the display and only turns black once the balance is deemed stable. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
The weighing display appears with the changed settings. This setting remains until the selection is changed. There is an online animation for this application. Click here to view it: Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 42
Some weight units and accuracy settings may be blocked from use in verified bal- ances, depending on national laws regarding legal metrology and verification. Depending on the country-specific model version, not all weight units listed may be available. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
The weight value is displayed in gray if it is smaller than the specified SQmin value. The weight value cannot be saved or printed out in applications 0.10370 (depending on the setting in the “Safety level” menu). SQmin 0.12000 g Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
This ID is queried before each printout. Requirements: – The balance must be connected to a Sartorius lab printer or PC for printouts (see page 78). – The settings for individual identifiers can be configured in the system settings...
Touch to confirm. Select to zero the balance if necessary. Place a container on the weighing pan. Select to tare the balance. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 46
= 0.00001 g Net1 0.00000 Component Place the next component on the weighing pan. To save the component, select Continue this procedure with additional formula components. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 47
The total weight is displayed and the application switches back to its original state. Printing Results Requirements: For printouts, a printer (e.g., Sartorius YDP40) or PC must be connected and configured (see page 78). Mixing Print component In the...
Determine whether the balance is tared when a component is saved. Touch to confirm. Select to zero the balance if necessary. Place a container on the weighing pan. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 49
Comp2 + 8.47000 g Comp3 + 18.30000 g Comp4 + 13.19000 g Comp5 + 13.18000 g To display the intermediate result for previously saved components, select Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 50
To exit the function, select y The application returns to its original state. There is an online animation for this application. Click here to view it: Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Touch to confirm. Select to zero the balance if necessary. Place the container for the first component on the weighing pan. Select to tare the balance. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 52
Place the second component on the weighing pan. To save the component, select y The weight value is displayed. Max 120 g d = 0.00001 g 32.59000 Component Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 53
The application returns to its original state. Printing Results Requirements: For printouts, a Sartorius YDP40 printer or PC must be connected and configured (see page 78). y If a printer is connected, an additional button automatically appears at the top right of the display.
= (W / (W )) * Requirements: A Sartorius density determination kit is required for this function (see Chapter “Accessories” on page 103). Mount the density determination kit on the balance and prepare it as described in the kit instructions.
Page 55
Place the sample on the weighing pan above on the frame of the density determination kit. y The weighing display shows the step Weight in air Max 120 g d = 0.00001 g 4.89090 Weight in air Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 56
If the GLP printout function is activated, the GLP printout is automatically printed (see page 33). There is an online animation for this application. Click here to view it: Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 57
0.78213 0.78205 0.78196 0.78187 0.78179 0.78170 0.78161 0.78153 0.78144 0.78136 0.78127 0.78118 0.78110 0.78101 0.78092 0.78084 0.78075 0.78066 0.78058 0.78049 0.78040 0.78032 0.78023 0.78014 0.78006 0.77997 Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Select the accuracy of the percentage display on the right. 100.0 0.00 Touch to confirm. 0.000 Select to zero the balance if necessary. Place a container on the weighing pan. Select to tare the balance. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 59
Place the unknown sample on the weighing pan. y The balance shows the percentage of the sample based on the reference sample. Max 120 g d = 0.00001 g 100.10000 Reference weight 14.32000 Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 60
To exit the function, select y The application returns to its original state. There is an online animation for this application. Click here to view it: Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
, enter the desired value, and confirm with Select the accuracy of the conversion display on the right. Touch to confirm. 0.00 Select to zero the balance if necessary. Select Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 62
To exit the function, select y The application returns to its original state. There is an online animation for this application. Click here to view it: Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Select to zero the balance if necessary. To weigh a living animal, place a container or cage on the weighing pan. Select to tare the balance. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 64
Or remove the sample from the weighing pan. y The application returns to its original state. There is an online animation for this application. Click here to view it: Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
The set thresholds are displayed. Max 120 g d = 0.00001 g 0.00000 90.00000 95.00000 Select to zero the balance if necessary. Place a container on the weighing pan. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 66
To exit the function, select y The application returns to its original state. There is an online animation for this application. Click here to view it: Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Without stability – setting is suitable for unstable samples. Touch to confirm. Select to zero the balance if necessary. Select Place the sample on the weighing pan. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 68
To exit the function, select y The application returns to its original state. There is an online animation for this application. Click here to view it: Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Touch to confirm. Select to zero the balance if necessary. Place a container on the weighing pan. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 70
The piece count is calculated and displayed. Place other pieces, if any, on the balance to reach a specific desired count. Max 120 g d = 0.00001 g Piece weight 0.18340 Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 71
To exit the function, select y The application returns to its original state. There is an online animation for this application. Click here to view it: Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Calibration/adjustment using an internal calibration weight – Calibration/adjustment using an external calibration weight (see “Calibration/Adjustment Using an External Calibration Weight” on page 74) – Automatic calibration/adjustment with isoCAL Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Adjustment deviation: Dev 0.00000 g y The balance is now calibrated and adjusted. Information on potential errors can be found in Chapter “Status Messages” on page 90. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Adjustment deviation: Dev 0.00000 g y The balance is now calibrated and adjusted. Information on potential errors can be found in Chapter “Status Messages” on page 90. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
The saved reports can be viewed via the system settings and printed out on a PC or Sartorius lab printer (see page 30). y If the ambient temperature has changed since the last calibration/adjustment or...
The header is included with the first printout. y To print the results of the application and the footer: Select the key. To print the footer: Select the key. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 77
Device 1234 A ID Inv.-Nr. 11 ---------------------- 2014-11-23 14:57 External calibration Start: manually 50.00000 g 3.00000 g External adjustment 0.00000 g ---------------------- 2014-11-23 14:57 Name: ---------------------- Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
9-pin (Printer YDP20-0CE) Sartorius order no.: (Second display YRD03Z) YCC03-D25 Serial device USB-RS 232 SBI with 25-pin USB-RS 232 xBPI Mini-A SUB-D plug RS 232 25-pin Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
PC with Windows 7, Vista, Windows XP, or 2000 operating system – A to Mini-B USB connection cable from PC to balance, Sartorius order no. YCC04-D09 In legal metrology data can be transmitted to a PC and used with an Alibi memory.
Page 80
Set the language to English (USA) – U.S. in the language toolbar of the PC (e.g., in the taskbar). Note: If English (UK) is chosen, “G£” is printed instead of “G#.” Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 81
Microsoft Excel 2010 (example): Gross/Tare/Net PC text format – With the setting on the balance, Microsoft Word 2010 has the following display for the “Gross/Tare/Net” example: Gross/Tare/Net Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 82
USB mass storage Go to and select the option. To connect the balance memory to the PC: Start Select USB mass storage Storage connect to PC? Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 83
Note: This does not apply to operating modes “PC spreadsheet format,” “PC text format,” and “Off.” The data transfer commands can be found in Chapter “Data Input Formats.” Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 84
Device (Slave) Mini-A Type-B Connect the balance to the Sartorius printer using the supplied USB connection cable. y The balance detects the printer automatically. No settings need to be changed. Connection for 9-pin RS-232 Configuration Printer YDP 20-0CE Sartorius order no.:...
Display values without a decimal point are printed without a decimal point. Normal Operation Space CR Carriage return Displayed characters Line feed Unit characters Decimal point or comma Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 86
In settings “PC - SBI” and “USB RS-232 SBI” the non-verified display digit is not automatically identified. Corresponding measures or settings must be carried out on the peripheral device. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
) In settings “PC - SBI” and “USB RS-232 SBI” the non-verified display digit is not automatically identified. Corresponding measures or settings must be carried out on the peripheral device. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Format 1 Format 2 Escape Command character Number Underscore (ASCII: 95) Carriage return (optional) Line feed (optional) Examples: Format 1: Esc P Format 2: Esc x1_ Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
ESC x2_ Print serial number ESC x3_ Print balance software version ESC x4_ Print software version of display and control unit ESC x5_ Print user/device ID Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
“Individual Identifiers”). “Not able to decrease When the sample ID could not be assigned a number in identifier.” descending order automatically (as described in Chapter “Individual Identifiers”). Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 91
“Unable to read file” When an error occurred when the file was being read. USB Device Error Message “The attached USB device When a USB device (printer) not made by Sartorius is is not supported.” connected. Secura, Quintix User Manual...
Transporting the Balance 12. Transporting the Balance Please use the original packaging for shipping. Packaging can be ordered through the Sartorius Service Center if required. Avoid glass breakage, shocks and vibrations: Never lift or carry the balance by its draft shield! Prior to shipping, switch the balance to standby mode (see page page 94) and then pull the power plug.
Place the cardboard in the recess provided for the purpose. Put the top part of the packaging over the device parts. Next, put the packaged balance into the cardboard box and seal the box. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
13.1 Service To ensure the continued accuracy of your balance, we recommend scheduling regular servicing at least once a year. The Sartorius Service Center offers different service contracts for this purpose, which can be individually adapted to suit your needs (see also www.sartorius.com/service).
Page 95
Push the side panels slightly backwards and remove them from the draft shield. Pull the front panel over the groove on the bottom of the draft shield and remove the panel. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 96
Reinsert the rear panel. Ensure that this engages in the locking mechanism above the rear panel. – Follow the instructions in Chapter “Installing the Balance” on page 10. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
2) Select the “Services” tab. 3) Then select “Disposal Information.” 4) Addresses for the local Sartorius disposal contacts can be found in the PDF files available for download on this page. Sartorius will not take back equipment contaminated with hazardous materials (ABC contamination) –...
** For verified balances in accordance with EU requirements, the legal regulations apply. Depending on the country-specific model version, not all weight units listed may be available. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 99
FCC radio frequency emission limits. Modifications: Any modifications made to this devices that are not approved by Sartorius Lab Instruments GmbH & Co. KG may void the authority granted to the user by the FCC to operate this equipment.
= SAR: Standard balances with country-specific additions for Argentina x = SJP: Standard balances with country-specific additions for Japan x = SKR: Standard balances with country-specific additions for South Korea Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
= OIN: Balance with Type Approval Certificate for India ) all Models ...CN: Type “SQP” Upper edge of the weighing pan to the lower edge of the upper draft shield panel Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 102
= OIN: Balance with Type Approval Certificate for India ) all Models ...CN: Type “SQP” Upper edge of the weighing pan to the lower edge of the upper draft shield panel Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Proof Line knob weight 50 g, OIML Class E2, with DAkkS certificate Calibration weight for lab balance type 35 YCW422-AC-02 Proof Line knob weight 20 g, OIML Class E2, with DAkkS certificate Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Visit www.scaleman.com your Authorized Sartorius Distributor Specifications 15.5 Balance Dimensions ® ® 15.5.1 Secura , Quintix 35, 65, 125, 125D, 225D, 324 All dimensions are given in millimeters Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 105
Visit www.scaleman.com your Authorized Sartorius Distributor ® 15.5.2 Secura All dimensions are given in millimeters Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 106
EU Declaration of Conformity Appendix The declaration of conformity supplied here is for verified balances for use in the EEA. Please keep it in a safe place. Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 107
Visit www.scaleman.com your Authorized Sartorius Distributor EC Type-approval Certificate Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 108
Visit www.scaleman.com your Authorized Sartorius Distributor Plates and Markings Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 109
Visit www.scaleman.com your Authorized Sartorius Distributor Specifications Secura, Quintix User Manual Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 110
Visit www.scaleman.com your Authorized Sartorius Distributor Visit www.balances.com for the full line of Sartorius Laboratory Balances...
Page 111
Visit www.scaleman.com your Authorized Sartorius Distributor Visit www.balances.com for the full line of Sartorius Laboratory Balances...