Unpacking, Assembly, and Safety Check
3. Adjust the F
from the gas source. Verify that e360 provides an audible alarm
and visual O2 Supply Loss alarm message.
4. Reconnect the high pressure oxygen hose to the gas source.
Verify that the alarm is no longer violated. Push Reset to clear
visual message and indicator. Return the F
5. Disconnect the high pressure air hose from the gas source. Verify
that e360 provides an audible alarm and visual Air Supply Loss
alarm message.
6. Reconnect the high pressure air hose to the gas source. Verify
that the alarm is no longer violated. Push Reset to clear visual
message and indicator.
AC Power Loss/Battery Backup Alarm
1. While the ventilator is operating, unplug the AC power cord
from the wall outlet. Verify that the ventilator continues functioning
and provides an audible and a visual alarm, the Int Battery
(Internal Battery) indicator lights and the message AC Power Loss
Battery Back Up appears in the window. The e360 issues a short
beep every five (5) minutes while running on internal battery.
2. Plug the AC power cord back into the wall outlet. Verify that the
ventilator continues functioning, the Internal Battery indicator goes
out and the Mains (Battery Charging) indicator lights.
3. Push Reset to clear visual messages and indicator.
High/ Low Airway Pressure Alarms / Circuit Disconnect Alarm/
Alarm Silence
1. Remove the test lung. Verify that both visual and audible
indicators for the Low Paw (low airway pressure) alarm are
activated after two mandatory breaths and after three breaths the
Circuit Disconnect alarm message is displayed.
2. Press the Alarm Silence button and verify that the audible alarm
is muted but the alarm lamp continues to blink and the alarm
message remains displayed.
3. Press the Alarm Silence button again and verify that the audible
alarm resumes beeping.
4. Re-attach the test lung. Verify that the audible alarm stops and
the 360° alarm lamp is steadily lit (latched).
5. Press the Reset button to clear all visual alarm indicators (lamp
and messages).
6. Remove the test lung and occlude the patient wye connector
5-1
of the breathing circuit. Verify that both the visual and audible
indicators for High Paw (high airway pressure) alarm are
activated.
7. Re-attach the test lung to the wye connector. Verify that the
audible alarm stops and the alarm indicator is steadily lit.
3-10
O
to .23. Disconnect the high pressure oxygen hose
I
2
O
to .21.
I
2
OPR360U A0509
OPR360US H0711