Test Connections
Refer to the Class I Insulation connection diagram
below.
Plug the appliance mains plug into the HPAT600
front panel mains socket (1).
Testing in Manual Mode
Press SHIFT and INS I (CLASS I) to select the
Class I insulation test and display the diagram screen.
Press OK for the HPAT600 to display the test screen
and perform the Class I insulation measurement.
If 'FAIL' is displayed after the Class I insulation
measurement, refer to 3.12. (Earth Continuity/ Insulation
Failure)
Testing in Auto Mode
Press SHIFT and INS I (CLASS I) to select the
Class I insulation test and display the test screen.
The HPAT600 will perform the Class I insulation test.
If 'FAIL' is displayed after the Class I insulation
measurement, refer to 3.12. (Earth Continuity/ Insulation
Failure)
28
Class I Test
Connections
2.9 Menu
Pressing the MENU button will display the first page of
the MENU screen.
Set Pass Levels
Set Insulation Voltage
Set Auto/Manual
Scrolling down with the
button will display the second
page of the MENU screen.
Recall Memory
Delete All Memory
Reset to Defaults
Use the
button to exit the MENU screen.
2.10 Null of Test Probe Lead Resistance (Class I
Earth Continuity only)
The null function is used to compensate for the test
probe lead resistance when used for Class I earth
continuity measurements. This also allows for the use
of alternative test probes.
To null the test probe lead resistance press SHIFT and
NULL (LOAD) to display the NULL screen. Use
select Null or Abort.
If Abort is selected, press OK and the HPAT600 will
return to the SELECT screen.
If Null is selected, connect the test probe from the
4mm socket (2) to the Earth pin of the front panel
mains socket (1) (use the brass earth contact pin
supplied with the HPAT600). While connected, press
OK.
Refer to 2.11
Refer to 2.12
Refer to 2.13
Refer to 2.16
Refer to 2.19
Refer to 2.20
to
9