Page 2
For the safety of the operator and the system, a thorough understanding of this manual is necessary before commissioning, using or maintaining the product. SSISTANCE For customer assistance please contact Meriam directly either by phone at (216) 281-1100, or by e-mail at meriam@meriam.com. ® HART is a registered trademark of the HART Communication Foundation.
MFT 4000 / 4005 / 4010 User’s Manual TABLE OF CONTENTS Subject Page MFT 4000 Overview ...........................1 Keypad Overview............................1 Single Function Keys..........................1 Dual Function Keys ..........................2 Soft Keys ...............................3 Alphanumeric / Symbol Entry ........................3 Alternative Numeric Data Entry ......................4 Measure Mode............................4 Header-line Symbols ..........................4...
Page 4
Hazardous Area Use (Intrinsically Safe Option)..................21 MFT 4000 Series - Model Specific Features ....................22 Calibrate Key – MFT 4000, MFT 4005, MFT 4010................... 22 Procedure (Manual, From PC, Template) ................... 23 Device Output (Conventional mA / V, HART mA, HART PV, User Entry……………………………… 23...
Page 5
Quick Menu Key (HART support key) ...................... 71 DOF Memory Maintenance ........................72 MFT 4010 Device Specific & Generic HART Communications..............73 Documenting HART Configurations with Meriam DMS................73 Download Site Information – DOFs, Firmware, Utilities ................73 Who has access to the Download Site ....................73 PC Requirements for Download Operations ..................
MFTs and applicable restrictions. KEYPAD OVERVIEW The following is an overview of the MFT 4000 series keypad operations. The MFT keypad has three basic key types (the functions of each are described below): Single function keys, dual-function keys, and soft keys. Single function keys control a dedicated MFT function.
0 to (decimal point) to (negative sign) and back to 0 for entry purposes. Calibrate This key enters the MFT into calibration & documentation mode for local indicators, conventional transmitters, HART transmitters, Fieldbus transmitters, and mA loop calibrations. info@Meriam-Direct.com 1.888.475.5235...
-, . , 0, 1…9, and space until the desired value is displayed. Press the Next soft key to enter that value and advance to the next place in the input field. Repeat as necessary. Press the Done soft key to end the numeric entry session. info@Meriam-Direct.com 1.888.475.5235...
AC adapter (P/N 9B000007) is connected and powering the MFT BUSY text box replaces either power symbol when BUSY the microprocessor is busy executing requested tasks NOTE: Power icons may temporarily disappear during certain HART Communication operations info@Meriam-Direct.com 1.888.475.5235...
A “D” symbol appears in the top line of the LCD display when exponential Damping is activated. To change the time setting for exponential Damping refer to the MFT Configuration Settings section of this manual. info@Meriam-Direct.com 1.888.475.5235...
Poff Used to turn off the 24V dc supply feature of the VMA0055 module. Vrng Used to select the V dc measurement range best suited to the application. See the VMA0055 manual in the Appendix for more information. info@Meriam-Direct.com 1.888.475.5235...
All addresses on the loop will be displayed on the MFT’s screen. The address of interest can be selected from the list to launch HART communication with that device. Digital Poll can only be launched from this location in the handheld’s menu structure. info@Meriam-Direct.com 1.888.475.5235...
Displays the minimum span of the EJA / EJX unit. Max Span Displays the maximum span of the EJA / EJX unit. Displays the lower Range Limit of the EJA / EJX unit. Displays the upper Range Limit of the EJA / EJX unit. info@Meriam-Direct.com 1.888.475.5235...
Important note: All MFT units are shipped with the PV Prompt locked out with a lockout code of 3-2- 1. See the “Saving Configurations for Multivariable HART Devices” section of this manual for more information on PV Prompt. info@Meriam-Direct.com 1.888.475.5235...
Note: Open sensor bays on the MFT should be filled with dummy modules (P/N A36741) to protect the electrical connector and avoid contamination of or damage to the sensor bays. Dummy modules are shipped as needed to fill MFT bays. info@Meriam-Direct.com 1.888.475.5235...
To install or remove batteries, follow the above instructions. After battery maintenance is complete, replace door and tighten the captive door screw. Note: For Intrinsically Safe MFT models, replace batteries only with approved types. See the MFT Intrinsic Safety Control Document in the Appendix for approved battery details. info@Meriam-Direct.com 1.888.475.5235...
Media Compatibility - Pressure Sensors Differential (xDN) pressure sensors are for use with clean, dry, non-corrosive gases only. Differential (xDI), Gauge (xGI) and Absolute (xAI) pressure sensor modules are compatible with fluid and gas media compatible with 316 stainless steel. info@Meriam-Direct.com 1.888.475.5235...
If your process has a fluctuating or pulsating pressure, use of an external pressure source is recommended to ensure the accuracy of your measurements and calibration. info@Meriam-Direct.com 1.888.475.5235...
Meriam does not recommend using process pressure as the pressure source for DP pressure device calibration. Errors in measurement and calibration can easily creep in if the pressure source is unstable. In addition, xDN sensors have a maximum overpressure limit of 150 PSI when a three-valve equalizing manifold is used properly.
The following diagram shows the connection for both a vacuum and an absolute pressure measurement. If the process maintains a stable pressure, it can be used as the pressure source in some calibrations. Process compatibility and pressure ranges of the sensor modules need to be considered for each application. info@Meriam-Direct.com 1.888.475.5235...
Important: Use the transmitter’s working terminals to measure mA, not the TEST terminals found on some transmitters. Do not use digital communication terminals found on smart transmitters. These terminals will give incorrect current reading. Voltage Transducer The following diagram depicts the correct set-up for transducer calibrations. info@Meriam-Direct.com 1.888.475.5235...
MFT or modules. These standards should meet the accuracy requirements for your company or industry. Meriam follows the guidelines established by ANSI / NCSL Z540-1-1994 which requires that the primary standard be 4 times more accurate than the unit under test.
Page 24
If Yes was selected, the screen will verify the new calibration by displaying the date and module position number. Press the OK soft key. The MFT then returns to the Measure Mode. Choosing No opens the Cal Not Saved screen with options: Exit Cal Restart Cal Save Cal info@Meriam-Direct.com 1.888.475.5235...
MFT firmware improvements, HART Device Object Files (DOFs) and documenting operations. All MFT upgrades are available via the Meriam Download Site accessible from www.meriam.com. The MFT does not need to be returned to Meriam for firmware updates or for additional of DOF loading.
Note: Restrictions apply to the use of Intrinsically Safe MFT units in hazardous areas. Refer to the Intrinsically Safe Control Document in the Appendix for exact details. Meriam reserves the right to revise the Control Document without notice. Contact Meriam for the current Intrinsically Safe Control Document.
CALIBRATE KEY – MFT 4000, 4005, & 4010 The Calibrate key provides access to powerful calibration & documentation tools. Pressing the Calibrate key allows the user to set up the MFT to calibrate / document calibration results of HART or non-HART devices.
Note The “From PC” calibration option requires the MFT Documentation option and Device Management System from Meriam or other compatible asset management software. To select a calibration type, move the selection arrow to Procedure and press Select. The following screen...
Loop Test: Loop testing of 2-wire HART or conventional device loops (see the Loop Test section of this manual for more details) Press the Back soft key if you choose not to make any changes. The MFT will then return to the Calibrate key main menu screen. info@Meriam-Direct.com 1.888.475.5235...
Page 30
Select the Function menu option under the Measure line to select the needed MFT operation for the device under test’s output. The following screen is shown: Meas: IV Meas P Meas TC Meas RTD Meas VMA Meas IV Meas Down Select Back info@Meriam-Direct.com 1.888.475.5235...
Once completed, the user can simulate the LRV and / or URV and simultaneously view the current output of the transmitter. This provides the user with the information necessary to determine what Smart Trims need to be run. See the RIO4000 and TIO0110 sections in the Appendix for more information on Simulation Modes. info@Meriam-Direct.com 1.888.475.5235...
Calibrate key main menu screen. 20 mA Analog Trim: MFT 4005 and 4010 execute 20 mA trim on HART devices. To perform the 4mA Trim, make sure the selection arrow points to 20 mA Analog Trim and press Select. info@Meriam-Direct.com 1.888.475.5235...
Press the OK soft key to continue. Press the Back soft key to return to the Hart Smart Trim Screen main menu screen. Pressing Back again will return the MFT to the Calibration key main menu screen. info@Meriam-Direct.com 1.888.475.5235...
The MFT will now prompt: Apply High Pressure. Press the OK soft key to continue. The MFT will then prompt the user: Wait until stable. This notice reminds the user to wait for the pressure to stabilize before accepting a pressure value. Press the OK soft key to continue. The following screen will appear: info@Meriam-Direct.com 1.888.475.5235...
Select Cal to view: PT-102 PT-307 DPT-101 DPT-106 TT-621 Down Select Back Move the selection arrow to the tag number desired and press the Select soft key. Point one of the saved calibration is displayed as shown below. info@Meriam-Direct.com 1.888.475.5235...
Use the Del soft key to delete the entire calibration record. Note: This Del key only removes the record from viewing by the MFT; it does not clear memory space. Connect the MFT to Meriam DMS or PC utility (pending) and use the clear MFT calibration data function to clear memory.
DMS, please go to www.meriam.com or contact Meriam at (216) 281-1100. The MFT 4000, MFT 4005, and MFT 4010 include documenting to enable the user to document Manual Calibration Procedures, From PC (Stored) Calibration Procedures and Template Calibration Procedures.
MFT operation for the device under test’s input. Measure menu allows the user to select the needed MFT operation for the device under test’s output. See the Calibrate Key section earlier in this manual for more information. A sample of the Cal Type screen is shown below. info@Meriam-Direct.com 1.888.475.5235...
“URV must be > LRV” will appear and the cursor will be pointed to the URV menu line. If the user pointed the Meas/Sim menu to an RIO4000 module for a Cal Type “P”, the message “Module / Type Mismatch!” info@Meriam-Direct.com 1.888.475.5235...
Page 40
2) start a simulation procedure using the simulation module selected. The Create Procedure set up screen is shown below. To change values simply move the cursor to the desired menu option and press the Select soft key. Use the MFT’s user interface to change or input the values desired. info@Meriam-Direct.com 1.888.475.5235...
Page 41
Create Procedure will cause the simulation module to automatically advance to the next point of the procedure upon saving of the previous point. Once the As found portion of a Manual Cal or Manual Procedure Cal is complete, the following display will appear: info@Meriam-Direct.com 1.888.475.5235...
Page 42
MFT calibration data. Note: HART configuration data saved by the MFT cannot be passed by FCINTF. This data can only be uploaded to Meriam DMS. To view previously stored calibration results, select the View Results menu option under the MFTs Calibrate key.
Continue until the As Found calibration procedure is complete Press the Done key to complete the “As Found” portion of the calibration and the As Found Complete Screen will appear (see below). info@Meriam-Direct.com 1.888.475.5235...
As Found Complete Screen From PC (Stored) Calibration Procedures and Documentation The Stored procedure option is only available when the MFT is used in conjunction with Meriam DMS software or other compatible FCINTF software programs. The Field Calibrator Interface (FCINTF) specification has been implemented as the MFTs Interface for receiving and transmitting calibration data.
Page 45
Apply the next input signal indicated by the Apply value and press the Save soft key when satisfied. Repeat as needed to complete the calibration. Press the Done soft key to end the As found portion of the calibration. The following display will appear: info@Meriam-Direct.com 1.888.475.5235...
Page 46
MFTs alphanumeric entry features. View Results – View saved calibration results from As left work. Returns to Cal Options Screen when finished Press the Done soft key to end the From PC (Stored) Procedure Calibration documentation session. info@Meriam-Direct.com 1.888.475.5235...
MFT tracks the Pass or Fail of each point in the procedure and marks the overall As found results and As left results as either Pass or Fail as appropriate. See the Meriam DMS User’s Manual for more details on Pass / Fail and the tag data entered in DMS to support this feature.
Once Cal Type is set to Loop Test then VMA Sim / IV Meas should automatically be set as well. When satisfied with the set up, Select the Begin Calibration menu option. On the resulting screen, shown at left below, info@Meriam-Direct.com 1.888.475.5235...
Page 49
Answer Yes to save the loop tag, confirm the information on Begin Calibration Setup Screen is correct, and then Select the Begin Calibration menu line when satisfied. If 1 pt, 2 pt, 3 pt or 5 pt Loop Test was selected then the Loop Test will launch as follows. info@Meriam-Direct.com 1.888.475.5235...
Page 50
The screen will update as shown below for a 2 point loop test. Complete the second point of the test in the same way. Loop Test Set: 20.000 mA Entered: 20.000 mA Error: 0.000 % 2 points of 2 Save Entry Abrt info@Meriam-Direct.com 1.888.475.5235...
Loop calibration procedure setup features similar to those described in the Loop Calibrations and Documentation section above are available in the Meriam Device Management System. This allows the user to define loop calibrations and assign calibration procedures to specific loop tag numbers in the DMS database.
Page 52
Cal Setup: Begin Calibration Procedure: Manual DeviceOut: Conv Cal Type: Switch RTD Meas / IV Meas Quick Cal Smart Trim View Results Clear Cal Memory Down Select Back Calibrate Key Main Menu Screen for Wet Bulb Temperature Switch info@Meriam-Direct.com 1.888.475.5235...
Page 53
LRV : 0.000 Reset Pt: 6.5.000 URV : 10.000 Direction : Inc Test Reset Pt : Yes Ramp Minutes: Start Delay Contact Form : NO VMA (wetting) : <More> Down Select Back Down Select Back Page 1 Page 2 (<More>) info@Meriam-Direct.com 1.888.475.5235...
Page 54
As Found data to As Left data, to Trim the switch set point (MFT display will show switch input and switch state while trimming operations are executed), to start the As Left documenting session (after Trim operations are completed) or to View the As Found Results. info@Meriam-Direct.com 1.888.475.5235...
An alternate way to clear documentation memory is to connect the MFT to Meriam DMS and use the Clear MFT Calibration Data function. See the DMS User’s Manual for more details.
The MFT 4005 provides the same calibration functions, NIST traceable module support and documenting functions as the MFT 4000 calibrator. In addition, all “Calibrate” key HART functionality present in the VAO1-MFT 4010 HART calibrator is also available (see the “Calibrate Key” section of this manual for more information). This includes the Quick Cal and Smart Trim calibration features that enable the user to fully trim HART pressure and temperature transmitters.
HART connections are made using the two standard banana jacks (3/4” center) located at the bottom of the MFT 4010. Refer to the following diagram. Polarity is not a concern for HART connections so both jack collars are black. Meriam supplies a test lead kit (P/N A36744) complete with standard connector for each model MFT 4010.
Poll may be executed to determine if the connected device has a non-zero address. Select Online Setup from the MFC Main navigation screen and then select the Multi soft key to initiate the Multidrop poll. If the “No Devices Found” message is received, then the loop connections to the MFC should be carefully checked. info@Meriam-Direct.com 1.888.475.5235...
If communication is still not established, check the List / Show DOFs menu under Offline Menu. Check the list for the presence of the “Generic” DOF under the manufacturer name “Meriam.” If “Generic” is found, then HART communication is possible when the device is properly wired and connected to the MFC (see connection troubleshooting details above).
Lists can be view by Manufacturer, by Device Name or by stepping through all Devices one at a time. Use the soft key controls to move through the list. Pressing the FUNCTION key changes soft key definitions Down and Up to PgUp and PgDn. To quickly advance to the next page of a list, press info@Meriam-Direct.com 1.888.475.5235...
Entries in this list may have been saved directly from a connected HART device, created in the MFT using the Create Configs function, copied from another configuration and renamed, or created by editing an existing file. MFT will also list configurations downloaded from the separately available Meriam Device Management System software (see www.meriam.com for more information).
This option allows the user to manage and configure sensor range and perform trim functions. From the Device Config screen scroll the cursor to Sensor and press the Select soft key. The following screen will appear: Sensor Menu 1151: Pump 8 Rerange Signal Conditioning Pressure Trim Down Select Back info@Meriam-Direct.com 1.888.475.5235...
Page 63
Typically, this value corresponds to the 20 mA output. The values entered must be within the span limits of the device. Move the cursor to URV and press the Select soft key. Rerange Menu 1151: Pump 8 Upper Range Value _150.00 InH20 Next Done info@Meriam-Direct.com 1.888.475.5235...
Page 64
Pres: 0.00 InH20 Down Select Back With pressure connections teed with to MFT and the HART device, apply the desired pressure and press the Select soft key. The MFT will then configure the pressure applied to equal 20 mA. info@Meriam-Direct.com 1.888.475.5235...
Page 65
Done. A prompt will appear: Save new data? Pressing No will return the MFT to the Signal Conditioning main screen without making any changes. Press Yes to accept the new value. The MFT _0.25 will return to the Signal Conditioning main screen. Next Done info@Meriam-Direct.com 1.888.475.5235...
Press the OK soft key to continue. Trim Zero Menu 1151: Pump 8 Pres: 0.00 InH20 Down Select Back Once the display is stabilized, press the Select soft key to set the displayed value. info@Meriam-Direct.com 1.888.475.5235...
Page 67
Pressing No returns the MFT to the Analog Trim main menu screen without saving changes. The screen will display the prompt: Remove Pressure. This notice reminds the user to release the pressure applied. Press OK to continue. The screen will display the prompt: Put Loop back online. This notice info@Meriam-Direct.com 1.888.475.5235...
Screen will prompt: Connect Ref meter. Press OK to continue. Press the Abrt soft key to return to the Analog Trim main menu screen without making changes. Analog Trim Menu 1151: Pump 8 Setting fld dev output to 4mA → OK Down Select Abrt info@Meriam-Direct.com 1.888.475.5235...
Select Scaled D/A Trim Display Scale Options & Corresponding I/O Resistor Values 1.0 to 5.0 volts corresponds to 250 ohm I/O resistor 1.2 to 6.0 volts corresponds to 300 ohm I/O resistor 2.0 to 10.0 volts corresponds to 500 ohm I/O resistor info@Meriam-Direct.com 1.888.475.5235...
Page 70
Selecting “Change scale” gives the user a chance to further adjust the desired scale. Once “Apply above scale” is selected, the remaining procedure is identical to the 1.2 to 6.0 volts example given above. Complete the procedure as prompted by the MFT displays using the custom scale for the D/A trim. info@Meriam-Direct.com 1.888.475.5235...
The screen will display the prompt: Loop may be returned to automatic control. This notice reminds the user to return the device to service. Press the OK soft key to continue. Press the Back soft key to return to info@Meriam-Direct.com 1.888.475.5235...
Message: Allows the user to enter a short message, similar to the Descriptor option above. Up to 20 characters, including spaces, may be used. Model: Provides information on the specific model of the connected HART device. This option is non- editable. info@Meriam-Direct.com 1.888.475.5235...
Move the cursor to Polling Address and press the Select soft key. The screen will display the current Polling address. Press the Edit soft key to change the current value, or press Back to return to the Hart Output main menu. After pressing Edit, the following screen will appear: info@Meriam-Direct.com 1.888.475.5235...
DIAGNOSTICS To enter the Diagnostics option, use the Up or Down soft key to move the selection arrow to Diagnostics, then press the Select soft key. The MFT screen provides two test options as shown below: info@Meriam-Direct.com 1.888.475.5235...
Page 75
MFT resets the PV to the original variable. Send: Allows the MFT to send (download) a HART configuration to an identical HART device. Once a suitable configuration for one unit is created, it can be shared with identical units requiring the same parameters. info@Meriam-Direct.com 1.888.475.5235...
HART Output: Refer to HART Output section for more information. An additional function of the Quick Menu key is to enter HART Offline mode. Press and hold the Quick Menu key for a few seconds to enter the Offline mode. info@Meriam-Direct.com 1.888.475.5235...
“Auto” button. An Auto Install dialog box will appear. Select the “Yes” response to the “DOF Update Options? question to erase all existing DOFs and re-load DOFs from the default DOF directory. 320 DOFs will take approximately 90 minutes to load. info@Meriam-Direct.com 1.888.475.5235...
DOWNLOAD SITE INFORMATION – DOFS, FIRMWARE, UTILITIES One of the unique features of the MFT 4000 Series is the built-in ability to have firmware upgrades installed by the user at his location. The Meriam Download Site is accessible over the Internet and provides for the convenient download of new or revised firmware and DOF’s (Device Object Files) for HART communication.
It should be installed onto the host PC using the DPC Manager CD included with shipment of the handheld, or from the Meriam Download Site, where it can be found at the bottom of the File Download Area page. The DPC Manager Utility coordinates download operations from the host PC to the MFT. Periodic improvements may be incorporated into DPC Manager.
RETURNING THE MFT FOR REPAIR In the event that a MFT requires service and must be returned to the factory for repair, please contact your local Meriam Representative (see list and contact information at www.meriam.com or the factory direct at the numbers listed below.
DAMPING – user selected options include no damping, Smart Damping, and exponential damping (user selectable time constant from 0.5 to 30 seconds to damp pulsation and stabilize the display). DIAGNOSTICS – MFT 4000 self-tests upon power up and upon installation of sensor modules. Advisories are posted on the display.
Page 82
1, 3, 5, or 9 points (exception: VMA0055 module must be returned to factory for recalibration). SECURITY LOCKOUT – MFT 4000 Settings can be locked out entirely or by specific feature. CE compliance info@Meriam-Direct.com...
MODELS MFT 4000 Multi-Function Tester – Base unit for use with up to three (3) sensor modules. ® ® MFT 4005 Multi-Function Tester with HART Trims – Base unit for use with sensor modules plus HART trims for pressure and temperature transmitters ®...
Page 84
4010 Calibrator with HART Calibration and Configuration Functionality Code OEM Field (or Reserved for outside price sheets) Add "MFT" as Prefix: Meriam Version Code Documenting Options Standard Documenting (Interface to DMS Software for Calibration or HART Configuration Data) 21CFR Part 11 Enabled Version of Documenting Option Above...
RTD mating connector (for use with customer's RTD) Simulate Cable; module connector by two (2) aligator clips For a precise list of Intrinsic Safety certified modules, please refer to the Intrinsic Safety Control Drawing in the Appendix of this manual. info@Meriam-Direct.com 1.888.475.5235...
Volt output the twelve (12) t/c types. Connections for measurement applications are made using mini-t/c connectors with appropriate t/c wire. Connections for simulation applications are made using mini-t/c connectors with wire cable and alligator clip ends (see Accessories section for the Meriam wire-kit part numbers). Measure Simulate •...
Page 88
Start Units: C Low pt: 100.0 Upr pt: 400.0 Steps: 3 Dwell Time: Manual Down Select Back 2.4.1 Select Start menu option to begin the indicated Auto-Step number. 2.4.2 Select Units to change the Auto-Step x simulation units. info@Meriam-Direct.com 1.888.475.5235...
Page 89
Refer to the MFT TIO0110 Calibration Manual for details of the re-calibration process. TC Setup User Re-Calibration Offsets CJ gain TC neg gain TC pos gain CJ Correction DAC coarse gain DAC fine gain DAC offset Full Recal Down Select Back info@Meriam-Direct.com 1.888.475.5235...
Page 90
T/C Wire Kit 1 for types E, J, K, N, T - mini t/c connector, 1 m cable, wire ends P/N A36858 T/C Wire Kit 2 for types B, K, R - mini t/c connector, 1 m cable, wire ends P/N A36881 mV simulation cable assembly; mini t/c connector, 3.5 ft. copper cable, alligator clips info@Meriam-Direct.com 1.888.475.5235...
Press Back when satisfied with the range selection. 1.2 For current measurements up to ±100 mA, use the VMA module’s “mAmp” and “Com” jacks as you would with a standard meter. ±100.000 mA can be measured without any range selection for current. info@Meriam-Direct.com 1.888.475.5235...
Page 92
The SX sensor location (depending which bay the VMA is located in) should read approximately 24.00 VDC. The MFT’s IV value will read the loop current. To establish HART communications with the test device, press the MFT’s HART key. info@Meriam-Direct.com 1.888.475.5235...
Page 93
3.4 Press the Setup soft key option for the VMA Current Settings: module to display the VMA Main Setup Display Start Mode mA Source Auto-Step 1 Auto-Step 2 Auto-Step 3 Step Value 1.000 mA Down Select Back VMA Main Setup Display info@Meriam-Direct.com 1.888.475.5235...
Page 94
4.1 Refer to Diagrams 3.1 and 3.2 and make connections for the desired application. 4.2 From the VMA Main Setup Menu, arrow down and select Auto-Step X. info@Meriam-Direct.com 1.888.475.5235...
Page 95
± (0.01% of reading + 0.015% FS) Open Circuit Voltage 24 VDC Output Drive 15 VDC minimum @ 24 mA, Resistive load Notes: 1. Nominal resistance at VMA current terminal is 10 – 15 Ω 2. Output load line is linear info@Meriam-Direct.com 1.888.475.5235...
Page 96
6.1 Intrinsically Safe by MET Laboratories, Inc per CAN/CSA-C22.2 No. 157, CAN/CSA-C22.2 No. 1010.1-92 & UL 913 Fifth Edition, Rev 2/21/97 for Class I, Division I, Groups A, B, C &D Hazardous Areas. Restrictions apply – see the Intrinsically Safe Control Document in Appendix for more details. 6.2 CE mark info@Meriam-Direct.com 1.888.475.5235...
RIO4000 Module Instructions For RTD (Resistance Temperature Detector) Measurement and Simulation when used with the MFT 4000 Series Multifunction Calibrator / HART Communicator. NOTICE: The calibrator firmware must be Version 3.42 or later in order for proper RIO4000 module operation.
Page 98
RIO4000 connection required to perform RTD Simulation for 2, 3 and 4 Wire RTD’s. Example 3-Wire RTD Example 2-Wire RTD Example 4-Wire RTD Configuration. Configuration. Configuration. Sensor Connections RIO4000 RIO4000 RIO4000 JUMPER JUMPER JUMPER RIO4000 Connections info@Meriam-Direct.com 1.888.475.5235...
Page 99
Fixed Simulation Value Entry Screen 1.2.4 Auto Step 1-3 menu options enable pre-programmed Simulations to be set up and executed automatically. Auto-Steps settings are remembered and stored for future use. Select Start menu option to begin the indicated Auto Step number. info@Meriam-Direct.com 1.888.475.5235...
Page 100
Rs – Probe ID Resistor (see Table below) 3, 4 One side of RTD 5, 6 Other side of RTD Pins 3 and 6 must be connected. See RTD wiring configuration options later in this document for more detail. info@Meriam-Direct.com 1.888.475.5235...
Page 101
4-wire RTDs have two wires/terminals of one color and two of another color. Connect pins 3 and 4 to each of two wires/terminals of the same color. Connect pins 5 and 6 to each of two wires/terminals of the other color. info@Meriam-Direct.com 1.888.475.5235...
Page 102
Install the RIO4000 module in one of the MFT bays and turn on the MFT. If no RTD is connected to the RIO module, the corresponding Sx line on the display will indicate “No Probe.” Be sure the RTD type is correct - if not correct, go to the RTD Sensor Module Setup display to change the type. info@Meriam-Direct.com 1.888.475.5235...
Page 103
Pt 500 (392) Ni 120 (672) Pt 500 (391) Fe 604 (518) Pt 500 (385) Down Select Back Down Select Back Down Select Back RIO User Select screen 2 RIO User Select screen 3 RIO User Select screen 1 info@Meriam-Direct.com 1.888.475.5235...
Page 104
Field Recal Enter Type of Cal: Field Recal Setup: Module Only Start Module + Probe Module Bay: 1 Units: C Points: 3 Restore Defaults Down Start Back Down Select Back Rcal Feature Main Setup Menu Calibration Type Selection Menu info@Meriam-Direct.com 1.888.475.5235...
Page 105
If the Cal Res value does not need to be edited, the user may move the cursor down to the Calibrate option on the Module Only Recalibration Screen and press Select . The calibration is completed and the following display is shown as confirmation. RTD Calibrate Calibration complete Date: 2004-05-20 Remove Cal Resistor Module Only Calibration Confirmation screen info@Meriam-Direct.com 1.888.475.5235...
Page 106
Press the Done soft key and then select Yes to save the new data point. This process adjusts that particular point on the calibration curve. Confirm the Error value is now within specification (see below). Field Recal Cal Point: Applied: -20.00 C -20.01 C Error: -0.01 C Abort Next Back info@Meriam-Direct.com 1.888.475.5235...
Page 107
Back: In the Rca l mode, Back aborts the calibration procedure and returns the user to the Rcal set-up screen. The previous calibration data is retained. It does not return the user to the previous screen in the calibration procedure. info@Meriam-Direct.com 1.888.475.5235...
Page 108
Measure cable assembly; RTD connector, 5 ft. copper cable, four (4) alligator clips P/N A900028-90201 Pt 100 RTD, DIN Class "A", 0.00385 alpha, 4 leads, 1/4" O.D., SS probe, 10" probe length, 3.5" handle and 5 ft coiled cable with connector for RIO4000 temperature module info@Meriam-Direct.com 1.888.475.5235...
Page 109
-60 to 160 Ni 120 -80 to 260 Fe 604 -100 to 200 Fe 908 -100 to 200 Fe 1816 -100 to 200 *Temperature Coefficient = 0.02 ˚C/˚C ** For 2-Wire Simulation, Manufacturer's 2 Wire Connector/Cable is Required. info@Meriam-Direct.com 1.888.475.5235...
Intrinsic Safety Document - Uncontrolled Multifunction Tester MFT 4000 Series For Controlled document see drawing sent with actual handheld info@Meriam-Direct.com 1.888.475.5235...
Descriptor Message Device Revision Software Revision Private Label Revisions Final Assbly Number “HART Output” Hart Output Polling Address Burst Mode Select Burst Command Number Device ID Preambles Diagnostic Menu” Diagnostics Self Test Loop Test Save Send Review Edit info@Meriam-Direct.com 1.888.475.5235...
Page 112
Burst Mode Select Flange Size Burst Com Number Number of Remote Seals Device ID Remote Seal Fill Fluid Preambles Remote Seal Type Perform Test Self Test Diagnostics Loop Test Test Results Save Save / Send Send Review / Edit Variables info@Meriam-Direct.com 1.888.475.5235...
Page 113
Number of Remote Seals Remote Seal Fill Fluid Remote Seal Type Device ID Hart Output Polling Address Request Preambles Burst Mode Select Burst Command Number Self Test Self Test Diagnostics Status Display Loop Test Save Save Send Send Variables Review Edit info@Meriam-Direct.com 1.888.475.5235...
Page 114
Disp Func Cond Engr Disp LRV Engr Disp Engr Disp URV Engr Disp Point Write Protect Write Write Enable 10min Protect New Password Self Test Self Test Diagnostics Loop Status Display Save Save Send Send Review /Edit Variables info@Meriam-Direct.com 1.888.475.5235...