6. Error messages/malfunctions
If an error occurs during a measurement, the measurement is discontinued
and a corresponding error code is displayed (Example: Error No. 1).
Error No.
Possible cause(s)
Err 1
No pulse has been detected.
Err 2
Unnatural pressure impulses influence the measurement result. Reason: The arm was moved during the Measurement (Artefact).
Err 3
The inflation of the cuff takes too long. The cuff is not correctly seated.
The measured readings indicated an unacceptable difference between systolic and diastolic pressures. Take another measurement
ERR 5
following directions carefully. Contact you doctor if you continue to get unusual readings.
There were too many errors during the measurement in 3Check mode, making it impossible to obtain a final result.
ERR 6
Read through the checklist for performing reliable measurements and then repeat the measurement.
«LO»
The pulse is too low (less than 40 beats per minute). Repeat the measurement.
The pressure in the cuff is too high (over 300 mmHg) OR the pulse is too high (over 200 beats per minute). Relax for 5 minutes and
«HI»
repeat the measurement.*