50
User Guide
Connectivity
Network/LAN
RS232
USB
Gas connection
Gas inlet
Gas outlet
Basic specifications
Warm-up time
Measuring ranges
Ambient temperature
Ambient pressure
Measurement gas
Measurement gas flow
Measurement gas temperature
Sample pressure range
1
Warm-up time is shortened at temporary power interrupts.
2
The flow will be lower than 27 ml/min when sample pressure is lower than atmospheric pressure.
PBI-300042-B 10/2010
EN
Ethernet 10/100 mbit/s Base-T with DHCP client or
fixed IP
D-SUB 9 DTE interface (male connedtor)
Host, USB 2.0 Connector type A, max current 100mA
Device, USB 2.0 Connector type B
Sample hose:
Sample type/connector: Needle 0,8x40mm
Filter/Water trap:
Connection for hose (inner diameter Ø 3mm)
Operational after 10 min.
(60 minutes to full specifications)
0-100 % O
and 0-100% CO
2
Operational:
+0 to +45 °C, less than 95 %RH,
non condensing
Storage:
-20°C to +60°C, less than 95 %RH,
non condensing
Operational:
900 hPa to 1050 hPa
Inert gasses (O
, CO
2
2
non condensing.
27 to 40 ml/min. (Typical flow 34 ml/min.)
5 °C to 35 °C
Ambient +100 mbar -200mbar
COPYRIGHT ©
CheckMate 3
Length 60 cm, Ø0,5mm
or
SmartPen 0,8x9mm
0,45 µ
1
2
, N
or Ar), less than 95 %RH,
2
2
PBI-Dansensor A/S