Power Supply Checks; Low Frequency Calibration; 1/10 Scale Adjustment; 1 Volt Adjustment - HP 3400A Operating And Service Manual

Rms voltmeter
Hide thumbs Also See for 3400A:
Table of Contents

Advertisement

Model 3400A
c.
When
the pointer
is
exactly
at
zero, rotate
the adjusting
screw
slightly
counterclockwise
to
remove
tension
on
pointer suspension.
If
the
meter
pointer
moves
to the
left
during
this
adjustment, repeat steps b
and
c.
5-19.
POWER
SUPPLY
CHECKS.
5-20.
Power
supply voltage
and
ac
ripple
tolerances
are
listed in
Table
5-4.
Test
points
are
also
indi-
cated
in this table.
When
making
ripple voltage
measurements,
it
may
be
desirable
to
use
a bat-
tery
powered
A. C.
Voltmeter
(H-P
403B)
to
avoid
any undesirable ground
loop currents.
5-21.
LOW
FREQUENCY
CALIBRATION.
5-22.
AMPLIFIER
GAIN
ADJUSTMENT.
a.
Connect
test
setup
illustrated
in
Figure
5-1.
b.
Set
Model
3400
A RANGE
switch
to
0.
01
volt
position.
c.
Adjust Voltmeter Calibrator
for
0.
01
volt
rms,
400
Hz
output; set
DC
Voltmeter
to
read
1.
0.
d.
Remove
Model 3400A
top cover; adjust
R4
(CAL)
for
1.
volt
as
indicated
on
DC
Volt-
meter.
If
R4
(CAL)
does
not
have enough
range
to calibrate the dc output, the
value
of
R3
should be changed.
The
range
of
R3
is
from
820
ohms
to
2.
16
k ohms.
e.
Adjust
R6 (FULL-SCALE ADJUST)
for
Model
3400A
full-
scale
meter
indication.
Section
V
5-23.
1/10
SCALE ADJUSTMENT.
a.
b.
Connect
test
setup
illustrated
in
Figure
5-1;
omit
the
DC
Voltmeter.
Set
Model 3400A
RANGE
switch
to
0.
1
volt
position
and
adjust
Voltmeter
Calibrator for
0.
01
volt
rms,
400
Hz
output.
NOTE-
The
1/10
SCALE ADJUST
should
be
set slightly
low
(needle's width)
to
reduce
meter
(needle) offset
with shorted
input.
c.
Adjust
R7
(1/10
SCALE
ADJUST)
for
Model
3400A
1/10 scale
meter
indication.
Change
value
of
R8
if it
is
necessary
to
increase
the
adjustment range
of
R7.
The
range
of
R8
is
from
approximately 270
ohms
to
430
ohms.
See Table
5-7.
5-24.
1
VOLT
ADJUSTMENT.
a.
Connect
test
setup
illustrated
in
Figure
5-1;
omit
the
DC
Voltmeter.
b.
Set
Model 3400A
RANGE
switch
to
1
volt
position.
c.
Adjust voltmeter
calibrator
for
1.
volt
rms,
400
Hz
output.
d.
Remove
right side
cover
(INPUT
side)
and
adjust
R104
(1
V
ADJUST)
for
Model 3400A
full-scale
meter
indication.
Table
5-4.
Power
Supply
Checks
POWER
SUPPLY
TEST EQUIPMENT
AND
CHECK
POINT
DC
VOLTAGE
SPECIFICATIONS
REGULATION
(Vary Line
Voltage
Between
103.
5
and
126. 5 vac)
RIPPLE
SPECIFICATIONS
-17.5
Vdc
Connect
DC
Voltmeter,
AC
Voltmeter, or
Os-
cilloscope
between
violet
lead
on video
amplifier
(A4)
board
and
chassis ground.
-16. 8 to
18.2
Vdc
±0.
5 volt
from
nominal
reading
at
115
Vac
line.
400
/J.V
rms
or
1.
1
mV
p-p
h-75
Vdc
Connect
DC
Voltmeter,
AC
Voltmeter,
or
Oscil-
loscope
between red/wht/
blue
on video amplifier
A4
and chassis ground.
70.0
to
78.0
Vdc
±
1
volt
from
nominal
reading
at
115
Vac
line.
400
/iV
rms
or
1. 1
mV
p-p
-6. 3
Vdc
Connect
DC
Voltmeter,
AC
Voltmeter, or Oscil-
loscope
between
Q3
emitter (grey
lead)
and
chassis ground.
-5. 9
to
-6.
5
Vdc
±0.
1
volt
from
nominal
reading
at
115
Vac
line.
750
(iV
rms
or
2.
mV
p-p
5-5