Page 3
Instrument Illustration Reference pH/ORP Sensor Junction under (Replaceable) Glass pH Bulb Electrode pH Glass Electrode Conductivity Cell (Built-in Electrodes) Temperature pH/ORP Sensor Sensor Protective Cap TEST Value Preprogrammed variable conductivity/ TDS ratios USER mode for programming special temperature compensation factor and °C conductivity/TDS RATIO...
INTRODUCTION Thank you for selecting the feature-packed Ultrameter II™, one of the Myron L Company’s latest in an increasing line of instruments utilizing advanced microprocessor-based circuitry and SMT manufacturing processes. This circuitry makes the instrument extremely accurate, reliable and very easy to use.
FEATURES and SPECIFICATIONS A. Features • Superior resolution 4 digit LCD displays full 9999 µS/ppm. • Accuracy of BETTER than ± 1% of reading in a handheld instrument. • All electrodes are internal for maximum protection. • Improved 4 electrode sensor technology. •...
Cond/TDS Ratio Factor D. Warranty/Service The Myron L Ultrameter II, excluding the pH/ORP sensor (6P), has a Two (2) Year Limited Warranty. The pH/ORP sensor (6P) has a Six (6) Month Limited Warranty for materials and workmanship. If an instrument fails to operate properly, see Troubleshooting Chart, pg.
III. RULES of OPERATION A. Operation Using the instrument is simple: • Individual or multiple parameter readings may be obtained by filling individual sensors or entire cell cup area. • Rinse the conductivity cell or pH/ORP sensor (6P) well with test solution 3 times and refill.
After use, the pH/ORP sensor well must be refilled with Myron L Storage Solution, and the protective cap reinstalled securely (ref. Maintenance of the pH/ORP Sensor, pg. 9 and Cleaning Sensors, 2. pH/ORP, pg. 32).
B. Maintenance of the pH/ORP Sensor (6P) The sensor well must be kept wet with a solution. Before replacing the rubber cap, rinse and fill the sensor well with Myron L pH Storage Solution. If unavailable, use an almost saturated KCl solution, pH 4 buffer or a saturated solution of table salt and tap water (ref.
IMPORTANT: After use, fill pH/ORP sensor well with Myron L pH Sensor Storage Solution and replace protective cap. If Myron L pH Sensor Storage Solution is unavailable, use a strong KCl solution, a pH 4 buffer, or a saturated solution of table salt and tap water (ref.
IMPORTANT: After use, fill pH/ORP sensor well with Myron L pH Sensor Storage Solution and replace protective cap. If Myron L pH Sensor Storage Solution is unavailable, use a strong KCl solution, a pH 4 buffer, or a saturated solution of table salt and tap water (ref.
you wish to change the solution type. NaCl Press and hold User for 3 seconds to make “SEL” appear (see Figure 1). For demonstration Figure 1 purposes, all 4 solution types are shown simultaneously. Use the key to select type of solution desired (ref.
2. Disabling Temperature Compensation Select USER mode, see Procedure to Select a Solution, pg. 11. With “USER” selected, press . If the display does not show .00%/°C, hold long enough to bring the tempco to .00%/°C (see Figure 3). Press twice (3 times if in TDS mode).
VII. CALIBRATION A. Calibration Intervals Generally, calibration is recommended about once per month with Conductivity or TDS solutions. Calibration with pH solutions should be checked twice a month. Calibration of ORP is not necessary (ref. CALIBRATION INTERVALS, pg. 19). B. Rules for Calibration of the Ultrameter II 1.
measurement key cancels changes not yet accepted and exits calibration mode. Leaving pH after the 2nd buffer results in the same gain being entered in place of the 3rd buffer. 2. Calibration Limits There are calibration limits. A nominal “FAC” value is an ideal value stored by the factory.
Rinse conductivity cell three times with your standard. Refill conductivity cell with same standard. Press , then press twice in COND/three times in TDS. The “CAL” icon will appear on the display. Press to step the displayed value toward the standard’s value or hold a key down to scroll rapidly through the reading.
5. pH Calibration (6P) Important: Always “zero” your Ultrameter II with a pH 7 buffer solution before adjusting the gain with acid or base buffers, i.e., 4 and/or 10, etc. a. pH Zero Calibration (6P) Rinse sensor well 3 times with 7 buffer solution. Refill both sensor wells with 7 buffer solution.
Page 22
reduce your variation from factory setting by pressing Press to accept the new value. The pH Zero Calibration is now complete. You may continue with pH Gain Calibration or exit by pressing any measurement key. b. pH Gain Calibration (6P) Important: Always calibrate or verify your Ultrameter II with a pH 7 buffer solution before adjusting the gain with acid or base buffers, i.e., 4 and/or 10, etc.
Press to accept 3rd point of calibration, which completes the Calibration procedure. Fill sensor well with Myron L Storage Solution and replace protective cap. 6. ORP Calibration (6P) ORP electrodes rarely give false readings without problems in the reference electrode. For this reason, and because calibration solutions for ORP are highly reactive and potentially hazardous, your Ultrameter II has an electronic ORP calibration.
D. pH and ORP Practices to Maintain Calibration (6P) Keep the sensor wet with Myron L Storage Solution. Rinse away caustic solutions immediately after use. ORP calibration solutions are caustic, and ± 5% is considered very accurate.
MEMORY This feature allows up to 100 readings with their temperatures to be stored simultaneously for later recall. At the same time, the TIME and DATE are also recorded. To download the memory to a computer, (ref. uDock™ IR Data Port, pg. 30). MEMORY A.
Press and scroll to location #3. Press and HOLD to clear old record #3. Fill pH/ORP sensor well with sample. Press to measure sample and press to store reading in location #3. The next memory stored will go into location #8. To clear all records: After MEMORY pressing...
Press the to change the time. Press to accept the change (new time). B. Setting DATE Example shown in Figure 12, is in US format i.e. mo/dy/yr. NOTE: The default format is US. Date format may be changed (ref. Date Format “US and International (Int)”, pg.
Press the to change the day. Press to accept the change (new day) Figure 15 (see Figure 15). C. DATE FORMAT “US & International (Int)” Press Press repeatedly until the format is displayed (stored ALL, time, date, date format). readings, PC OFF, CL Press to change.
format location. The display will show a “C” or “F” (see Figures 18 and 19). Figure 18 Figure 19 Press to switch units. Press to accept unit preference for all temperature readings. NOTE: Tempco will still be shown in %/°C . XII.
XIII. CELL CHECK The cell check verifies the cleanliness of the conductivity/TDS/resistivity sensor. In normal use the cell may become dirty or coated and require cleaning. If the display is showing “.00” when the cell cup is dry, the sensor is probably clean. However, when testing high purity water in resistivity (RES) mode improved accuracy may be desired.
Press to display the stored memory records. Press repeatedly until you pass the CELL ch location. The display will show “Auto oFF” (see Figure 24). to initiate. CAL Press will be displayed along with the “15 SEC” (see Figure 25). Press Figure 25 change the time (see...
Verify/Set the calibration linc. (See below – Setting User Mode Calibration Linc). B. Setting User Mode Calibration “Linc” The linc function sets or “links” the calibration gain factor of a Standard Solution to the User solution mode. Once set, the “linc” will stay intact with future calibrations unless the linc has been canceled.
Press key to accept the setting. Pressing any of the measurement keys will exit without changing the setting. User mode “linc” is now complete. The User mode will now use the calibration gain constant used for the calibration of the Standard Solution as outlined above.
IR DATA PORT INSTRUCTIONS Requires Myron L uDock™ accessory package, Model # U2CIP. The Myron L uDock is powered via the USB port, requiring no external power source. The uDock application will operate on Windows 2000 & XP*, and Macintosh OS9.2 & OSX** based computer systems.
Page 35
Once downloaded, the data may be manipulated, printed or stored within the Myron L uDock application, or the data may be exported to another more powerful spreadsheet † such as Excel*. † Please Note: Although the Myron L Company has performed extensive testing, we can not guarantee compatibility of all applications and formats.
XVII. CARE and MAINTENANCE Ultrameter IIs should be rinsed with clean water after use. Solvents should be avoided. Shock damage from a fall may cause instrument failure. A. Temperature Extremes Solutions in excess of 71°C/160°F should not be placed in the cell cup area;...
Page 37
Leaving high pH (alkaline) solutions in contact with the pH sensor for long periods of time is harmful and will cause damage. Rinsing such liquids from the pH/ORP sensor well and refilling it with Myron L Storage Solution, a saturated KCl solution, pH 4 buffer, or a saturated solution of table salt and tap water, will extend the useful life.
XVIII. TROUBLESHOOTING CHART Symptom Possible Cause No display, even though Battery weak or not connected. measurement key pressed Inaccurate pH readings 1. pH calibration needed. (6P) Ref. pH Cal., pg. 17. 2. Cross-contamination from residual pH buffers or samples in sensor well.
14.0 mS is the recommended standard. Order NACL-14.0 B. pH Buffer Solutions (6P) pH buffers are available in pH values of 4, 7 and 10. Myron L Company buffer solutions are traceable to NIST certified pH references and are color-coded for instant identification. They are also mold inhibited and accurate to within ±...
D. Soft Protective Carry Cases Padded Nylon carrying case features a belt clip for hands-free mobility. Two colors to choose from; Blue - Model #: UCC Desert Tan - Model #: UCCDT E. Hard Protective Carry Cases Large case with 2 oz. bottles of calibration standard solutions (KCl-7000, 442-3000, 4, 7, &...
varies with concentration and temperature in a non-linear fashion. Other solutions have more variation still. The Ultrameter II uses corrections that change with concentration and temperature instead of single average values. See Chart 1. 2.500% 2.400% 2.300% 2.200% 2.100% % / °C 2.000% 1.900% 1.800%...
D. A Chart of Comparative Error In the range of 1000 µS, the error using KCl on a solution that should be compensated as NaCl or as 442, is illustrated in the graph below. 442 error with KCl tempco NaCl error with KCl tempco (1)% Temperature (2)%...
(anions) in freshwater that the Myron L Company calls “natural water”. These are modeled in a mixture called “442™” which the Myron L Company markets for use as a calibration standard, as it does standard KCl and NaCl solutions.
readings for a specific solution. That value remains a constant for all measurements, and should be reset for different dilutions or temperatures. C. When does it make a lot of difference? First, the accuracy of temperature compensation to 25°C determines the accuracy of any TDS conversion.
over this change of 10°C. The solution is then said to have a tempco of 2.2 %/°C. Tempco always varies among solutions because it is dependent on their individual ionization activity, temperature and concentration. This is why the Ultrameter II features mathematically generated models for known salt characteristics that also vary with concentration and temperature.
XXIII. pH and ORP (6P) A. pH (6P) 1. pH as an Indicator (6P) pH is the measurement of Acidity or Alkalinity of an aqueous solution. It is also stated as the Hydrogen Ion activity of a solution. pH measures the effective, not the total, acidity of a solution.
Page 48
Electrode wire in a layer, obstructing contact with KCl solution Figure 33 test solutions. See pH/ORP, pg. 43. 4. The Myron L Integral pH Sensor (6P) The sensor in the Ultrameter II (see Figure 34) is a single ions Glass...
“ground” voltage comes from a reference junction - same as the pH sensor uses. 4. The Myron L ORP Sensor (6P) Figure 34, pg. 44, shows the platinum button in a glass sleeve. The same reference is used for both the pH and the ORP sensors. Both pH and ORP will indicate 0 for a neutral solution.
It can be cleaned off with detergent or acid, as with the pH glass. XXIV. SOFTWARE VERSION Contact the Myron L Company to see if a software upgrade is available. Press key. Press key until three numbers are displayed as shown in Figure 35.
XXV. GLOSSARY Anions Negatively charged ions. See Solution Characteristics, pg. 40. Algorithm A procedure for solving a mathematical problem. See Temperature Compensation and TDS Derivation, pg. 41. Logarithm An arithmetic function. See pH Units, pg. 43. Oxidation-Reduction Potential or REDOX, See ORP/ Oxidation-Reduction Potential/REDOX, pg.
Page 56
Myron L COMPANY 2450 Impala Drive Carlsbad, CA 92010-7226 Tel: +1-760-438-2021 Fax: +1-760-931-9189 E-Mail: info@myronl.com techquestions@myronl.com www.myronl.com Made In USA UMIIOM 11NO08...
Need help?
Do you have a question about the ULTRAMETER II 6P and is the answer not in the manual?
Questions and answers