Page 2
Internal view of Steripower 10041 / 10046 10046 without socket > operation only possible with rechargeable battery! Available versions 10041 (with rechargeable battery) 10048 (with internal power supply unit + signalling) 10046 (high-frequency device) 10127 secure light 10047 (with internal power supply unit)
Internal view of Steripower art. 10047 (Steripower with power supply unit) OVERVIEW & DESCRIPTION Description of the components*: External housing Strap for housing opening Disinfection area Spray nozzle Female plug for charging device Rechargeable battery Plug for rechargeable battery (+/-)
TABLE OF CONTENTS INTRODUCTION This manual provides to the owner and user shall contribute to the reliability of the device. designated instructions regarding installati- The selection of the installation site, place- on, usage and maintenance of the disinfection ment and installation of the device(s) shall be device.
TRANSPORT & STORAGE TRANSPORT Use the original packaging when transporting the device and observe the international information and protection symbols on the packaging during transport and storage! Set the cardboard box and/or device in this direction only! 290 mm Risk of breakage, protect against damage! Protect against wet conditions! Do not stack more cardboard boxes than indicated! 200 mm...
> Information labels attached to the device which were made illegible must be requested and replaced immediately.. > Use the Steripower and its accessories according to the intended and proper use and as specified by the manufacturer. SAFETY RULES | 04...
5x = 2,5 ml 10x = 5,0 ml predominantly used by children; distance between the bottom of the Steripower and the ground). Should the liquid quantity be to low or too high, 4. Rechargeable battery connection: charging is fine tuning can be performed by turning a only possible with the cable correctly connected to watchmaker‘s screwdriver to the right or left.
Page 9
Rechargeable battery is ready for operation! • red-green simultaneously: med. Recharge battery! Only for Steripower with rechargeable battery (art. 10040) • flashing red: Recharge battery immediately! Only for Steri- power with rechargeable battery (art. 10040) 9.
OPERATION & OPERATING PRINCIPLE FIELDS OF APPLICATION The automatic spraying device Steripower® was specifically designed for hand hygiene purposes in order to ensure quick, simple and effective disinfection of the hands. Fields of application are, for example: hospitals, medical practices, reti-...
Page 11
OPERATION & OPERATING PRINCIPLE SPRAY & APPLICATION TIMES Recommended spray quantity in ml and application time in min: • hygienic hand disinfection: at least.: 3 ml (quantity) at least.: 5 ml (quantity) • surgical hand disinfection: The hands must be sprayed and/or kept humid according to the specified application time of the alcoholic hand disinfectant.
Page 12
OPERATION & OPERATING PRINCIPLE REPLACING THE LIQUID CONTAINER > Pull down the strap at the bottom of the cover. > Remove the front cover by slightly lifting and then pulling it off towards the front. > Remove the plug and pull the suction pipe out of the empty container. >...
Page 13
OPERATION & OPERATING PRINCIPLE GENERAL NOTES Before opening the hood, disconnect the Steripower from the power supply if the device is directly connected to the mains (art. no. 10047), e.g. by switching off the fuse! • Before performing maintenance works or in order to disconnect the device from mains opera- tion, the power supply must be disconnected by pulling the mains plug and/or switching off the fuse.
MAINTENANCE & CARE DISINFECTANTS Alcohol (ethanol, isopropanol, N propanol) with +/- 70% alcohol content has the best proven effect in disinfectants, i.e. all common alcoholic hand disinfectants with approval. MAINTENANCE & CARE Only have maintenance works performed on technical or electronic parts of the device by technical personnel or the building services department.
ELECTRONIC SYSTEM ELECTRONIC SYSTEM Listing of the components: Printed circuit board (electronics) Plug for LED ready lamp Pushbutton for liquid quantity regulation (0.5 - 5 ml) Plug for motor pump Plug for non-return valve Plug for rechargeable battery 12 V - 7 Ah Fuse 1.6 A (delayed-action fuse) LED green (for operation) LED red (signal)
TROUBLESHOOTING In the case of a malfunction of the device, do not send it to us for repairs if you have not contacted your authorised dealer or the service hotline. Troubleshooting is usually possi- ble by service telephone under +49 (0) 8151 / 555 15 27. QUESTIONS &...
Page 17
• The device is operated with the rechargeable battery (art. 10041). • The device can be operated with the rechargeable battery (art. 10041) and the charging device plugged into a socket. The rechargeable battery continues to operate the device even in the case of a power failure.
Operation manual lead charging device TECHNICAL DATA • For 12 V rechargeable lead (Pb) batteries • Electronic protection against reverse polarity • Electronic protection against overloading • Electronic protection against short circuit • Charging according to IUoU characteristic • Automatic additional charge - floating charge •...
Page 19
The device must only be cleaned with a dry cloth. Do not put the device into operation when the housing or the mains plug is damaged. In case of defects, please contact an authorised dealer or service@steripower.de SAFETY NOTE The charging device may only be operated in closed and dry rooms. In order to exclude the risk of fire or the risk of an electric shock, the device must be protected against humidity.
Page 20
Touchless Hand-Disinfection unit Steripower® / white edition / stainless steel edition Typenbezeichnung und ggf. Artikelnummer Type designation and, if applicable, article no.