The Conductivity Probe can be used to
measure either solution conductivity or
total ion concentration of aqueous samples
being investigated in the field or in the
laboratory. Conductivity is one of the most common environmental tests of aquatic
samples. Even though it does not tell you specific ions that are present, it quickly
determines the total concentration of ions in a sample.
Note: Vernier products are designed for educational use. Our products are not
designed nor are they recommended for any industrial, medical, or commercial
process such as life support, patient diagnosis, control of a manufacturing
process, or industrial testing of any kind.
What's Included
Conductivity Probe
90 mL bottle of 1000 µS/cm NaCl standard solution
Compatible Software and Interfaces
See
www.vernier.com/manuals/con-bta
compatible with the Conductivity Probe.
Getting Started
1. Connect the sensor to the interface (LabQuest Mini, LabQuest 2, etc.).
2. Start the appropriate data-collection software (Logger Pro, Logger Lite,
LabQuest App) if not already running, and choose New from File menu.
The software will identify the sensor and load a default data-collection setup. You
are now ready to collect data.
If you are collecting data using a Chromebook™, mobile device such as iPad
Android™ tablet, or a Vernier wireless sensor or interface, please see the following
link for up-to-date connection information:
www.vernier.com/start/con-bta
Using the Product
1. Set the range switch on the sensor.
2. Thoroughly rinse the lower section of the probe using distilled or deionized
water.
3. Connect the sensor following the steps in the Getting Started section of this
user manual.
4. When you are finished making measurements, rinse the electrode with distilled
water. Store dry.
for a list of interfaces and software
Videos
View videos related to this product at
Calibration
For many experiments, calibrating the Conductivity Probe is not required. A
calibration equation is stored on each probe before they are shipped, which is used
as a default by Vernier software.
For the most accurate measurements with this sensor, we recommend calibration. It
is a simple process that takes only a few minutes.
For instructions for Conductivity Probe calibration using Logger Pro computer
software, see
For instructions for Conductivity Probe calibration using LabQuest App, see
www.vernier.com/til/3394
For instructions for Conductivity Probe calibration using Graphical Analysis
with a Chromebook, see
For instructions for Conductivity Probe calibration using Graphical Analysis
with an iOS or Android device, see
In order to calibrate a Conductivity Probe, or to confirm that a saved calibration is
accurate, you should have a supply of conductivity standard solutions that cover
the range of the conductivity values you will be measuring. For more information
about conductivity standard solutions, including recipes for preparation, see
www.vernier.com/til/760

Specifications

Low range
Mid range
High range
Type
Response time
Temperature compensation automatic from 5 to 35°C
Temperature range
®
or
Accuracy using factory
calibration
Accuracy using custom
calibration
Shaft diameter
Care and Maintenance
When you have finished using the Conductivity Probe, simply rinse it off with
distilled water and blot it dry using a paper towel or lab wipe. The probe can
www.vernier.com/con-bta
www.vernier.com/til/2341
www.vernier.com/til/3631
www.vernier.com/til/3630
0 to 200 µS/cm (0 to 100 mg/L TDS)
0 to 2000 µS/cm (0 to 1000 mg/L TDS)
0 to 20,000 µS/cm (0 to 10,000 mg/L TDS)
ABS body, parallel graphite electrodes
98% of final reading in 5 seconds
0 to 80°C
±8% of full-scale reading for low range
±3% of full-scale reading for mid range
±4% of full-scale reading for high range
±2% of full-scale reading for each range
12 mm OD
2

Advertisement

loading
Need help?

Need help?

Do you have a question about the Conductivity Probe and is the answer not in the manual?

Questions and answers

Summary of Contents for Vernier Conductivity Probe

  • Page 1: Specifications

    In order to calibrate a Conductivity Probe, or to confirm that a saved calibration is accurate, you should have a supply of conductivity standard solutions that cover What's Included the range of the conductivity values you will be measuring.
  • Page 2: Troubleshooting

    How the Sensor Works The Vernier Conductivity Probe measures the ability of a solution to conduct an electric current between two electrodes. In solution, the current flows by ion transport. Therefore, an increasing concentration of ions in the solution will result in higher conductivity values.
  • Page 3 If you are taking readings at temperatures below 15°C or above 30°C, allow The salinity range of the Conductivity Probe is 0 to 10 ppt. Seawater has a salinity  more time for the temperature compensation to adjust and provide a stable of 35 ppt, so any seawater samples will need to be diluted before making conductivity reading.
  • Page 4 Accessories/Replacements Rev. 2/8/16 Item Order Code Logger Pro, Logger Lite, Vernier LabQuest 2, LabQuest Mini, and other marks shown are our trademarks or Conductivity Standard Solution (Low, 150 µS/cm), 500 mL CON-LST registered trademarks in the United States. Conductivity Standard Solution (Middle, 1413 µS/cm), 500 mL CON-MST All other marks not owned by us that appear herein are the property of their respective owners, who may or may not be affiliated with, connected to, or sponsored by us.