Sirona Dental Systems GmbH
Operating Instructions SINIUS / SINIUS CS
63 22 726 D3561
D3561.201.01.03.02 05.2012
4.11.2.3
Connecting the SIROCAM AF intraoral camera
Connecting the SIROCAM AF intraoral camera
Connecting the SINIUS intraoral camera to the dentist element
Depending on the equipment version, the intraoral camera can be placed
at instrument position five or in the additional holder.
The camera connector is connected to the lower side of the dentist
element. If the dentist element does not have an additional holder, the
camera connection at instrument position five will be located below the
dentist element A. If an additional holder is available, then it is connected
to an additional sixth position B.
➢ Plug the connector of the intraoral camera into the correspondingly
marked sockets on the dentist element.
The plug locks in place.
Connecting the SINIUS CS intraoral camera to the dentist element
The connecting cable of the intraoral camera is guided through a swivel
arm. The camera connection is located below the instrument holder. The
holder is fixed on the front edge with two internal clamps on the dentist
element.
1. The instruments must be removed from the instrument holder. To do
this, remove all instruments one after the other and allow these to
overhang in front of the dentist element.
2. Lift the instrument holder by the front edge until the clamps are
released and the holder can be removed.
3. The connection for the intraoral camera is located at the fifth
instrument position. Mount the connection cable of the camera in the
swivel arm and insert the camera connector into the socket.
The plug locks in place.
4. First insert the instrument holder into the groove on the dentist
element. Then push the holder forward and downward until it locks
into place.
5. Place the instruments in their holders. Make sure that the instrument
hoses and the connection cable of the camera are located in the
guide rolls of the swivel arm.
Connecting the intraoral camera to the assistant element (only for
SINIUS CS)
If the treatment center is equipped with the SINIUS CS dentist element,
the intraoral camera can be operated on the first instrument position of
the assistant element. The camera connection is now located below the
assistant element instead of the curing light connection.
➢ Plug the connector of the intraoral camera into the correspondingly
marked socket on the assistant element.
The plug locks in place.
4 Operation
4.11 SIVISION digital video system
147
Need help?
Do you have a question about the Sinius CS and is the answer not in the manual?
Questions and answers
WHY IS WATER LEAKING CONSTANTLY FROM UNDER NEATH BY SCALER ON BRACKET TABLE NOT FROM ATTACHMENTS IT FROM A METAL CIRCLE
The constant water leakage from underneath the scaler on the Sirona Sinius CS bracket table could be due to an issue with the water supply, drainage, or an internal malfunction in the scaler system. Possible causes include:
1. A defective water supply connection or leakage in the internal tubing.
2. A malfunction in the SIROSONIC TL scaler, possibly caused by improper operation or a faulty component.
3. A blockage or improper installation in the drainage system, leading to water accumulation and leakage.
To resolve this issue, check the water connections, ensure proper drainage, and inspect the scaler for defects. If the problem persists, inform your service engineer.
This answer is automatically generated