Adjustments
973A
972A
Calibration Table
CAUTION
Dangerous voltages are present during the calibration procedure. Calibration should only be
performed by qualified service technicians. Use a non-conductive tool.
Step
Function
1
2
mV
3
4
5
V
6
Range
Input Signal
Short
400 mV
380.0 mV
40 mV
38.00 mV
400 V
380.0 V
4 V
3.800 V
40 V
38.00 V
Tolerance (counts)
Adjustment
(limits)
972A
±1
—
±8
1 (±1)
±18
2 (±1)
±8
3 (±1)
±8
4 (±1)
±8
5 (±1)
973A
±1
±4
±16
±4
±4
±4
6 - 1