Adjustments; Calibration Table - HP 972a User Manual

Table of Contents

Advertisement

Adjustments

973A
972A

Calibration Table

CAUTION
Dangerous voltages are present during the calibration procedure. Calibration should only be
performed by qualified service technicians. Use a non-conductive tool.
Step
Function
1
2
mV
3
4
5
V
6
Range
Input Signal
Short
400 mV
380.0 mV
40 mV
38.00 mV
400 V
380.0 V
4 V
3.800 V
40 V
38.00 V
Tolerance (counts)
Adjustment
(limits)
972A
±1
±8
1 (±1)
±18
2 (±1)
±8
3 (±1)
±8
4 (±1)
±8
5 (±1)
973A
±1
±4
±16
±4
±4
±4
6 - 1