RETURN OF GOODS TO MANUFACTURER If equipment needs to be returned to Ludlum Measurements, Inc. for repair or calibration, please send to the address below. All shipments should include documentation containing return shipping address, customer name, telephone number, description of service requested, and all other necessary information.
Model 2200 Scaler Ratemeter Section Introduction The Model 2200 Scaler Ratemeter is a self-contained counting instrument designed for operation with scintillation, proportional, or GM (Geiger Muller) detectors. Counting information is displayed both as a scaler count on the LED (light-emitting diode) display and as an averaged count rate on the analog scaler.
Model 2200 Scaler Ratemeter Section Specifications 1 mA, 6.4 cm (2.5 in.) scale, self-shielded DC mechanism Meter: four-decade, linear ratemeter with ranges of 0-500; 0-5000; 0- Ratemeter: 50,000; and 0-500,000 cpm (counts per minute); 6.4 cm (2.5 in.) scale panel meter;...
Page 11
Model 2200 Scaler Ratemeter -20 to 50 °C (-4 to 122 °F); may be certified to operate Temperature Range: from -40 to 65 °C (-40 to 150 °F) 21.6 x 12.7 x 21.6 cm (8.5 x 5 x 8.5 in.) (H x W x L), excluding handle...
Model 2200 Scaler Ratemeter Section Description of Controls and Functions Front Panel COUNT LAMP: a red LED, indicating that the scaler is in the count cycle. COUNT SWITCH: resets and starts the scaler counting. The scaler turns off at the end of the preset time.
Model 2200 Scaler Ratemeter ZERO Switch: When depressed, this switch discharges the integrating capacitor, driving the meter to zero. F-S Response: a two-position toggle switch indicating fast or slow response. In “F” position, the meter will deflect from 10 to 90% of full scale in 4 seconds. In “S” potion, the meter will deflect from 10 to 90% of full scale in 22 seconds.
Pollution Degree 2 (as defined by IEC 644) Cleaning Instructions and Precautions The Model 2200 may be cleaned externally with a damp cloth, using only water as the wetting agent. Do not immerse the instrument in any liquid. Observe the following precautions when cleaning: 1.
The operator or responsible body is cautioned that the protection provided by the equipment may be impaired if the equipment is used in a manner not specified by Ludlum Measurements, Inc. Caution! Verify instrument voltage input rating before connecting to a power converter.
Model 2200 Scaler Ratemeter The “crossed-out wheelie bin” symbol notifies the consumer that the product is not to be mixed with unsorted municipal waste when discarding; each material must be separated. See section 8, “Recycling” for further information. Also displayed on the side panel.
Model 2200 Scaler Ratemeter Section Operating Procedures Power Select either line or battery operation with the power switch. Line Operation: Connect the instrument to line power of 95-250 Vac, 50-60 Hz. Turn the power switch to LINE and proceed to use the instrument.
Model 2200 Scaler Ratemeter Adjust DISCR until scaler meter reaches 75% of generated, incoming count rate. The input sensitivity is now set for 10 mV with THRESHOLD set for 1.00. Proceed to use instrument with THRESHOLD setting at 1.00 or more.
Model 2200 Scaler Ratemeter graph, relating count rate to system gain. This relationship is commonly referred to as a plateau or instrument plateau curve. System gain may be changed by adjusting detector high voltage (HV), THRESHOLD control or DISCR (amplifier gain).
Window Operation and Energy Calibration Procedures General The Model 2200 is calibrated at the factory so that one turn on the THRESHOLD controls is equal to one turn on the window dial. Equipment Required Detectors capable of energy discrimination. Examples used are the Ludlum Models 44-3 and 44-2.
55 keV and 65 keV. Adjust the THRESHOLD dial for other energies of interest. Procedures to Calibrate Threshold to 100 keV/Turn Initial Model 2200 control settings: Switch the window ON-OFF to OFF. Ratemeter RANGE selector switch at X10. THRESHOLD dial is at 6.42.
Section Data Output General The Model 2200 RS-232 serial port can be connected to a computer or printer for data-logging of the scaler count information. Windows -based software and cable are supplied with the instrument. The computer software can control the count time, start and stop counting, time/data stamp data, and print or save data.
Model 2200 Scaler Ratemeter 3. Prior to running the program, ensure that the Model 2200 is in the PC mode, and the count time push-wheels are set to 000. Button Functions Start/Stop Count: Click on this button to start a count. Clicking on it again will hold the count.
Model 2200 Scaler Ratemeter RS-232 Commands The Model 2200 communicates by way of the RS-232 interface. The RS-232 port operates at 2400 baud, 8 data bits, 1 stop bit, no parity, and hardware (RTS/CTS) flow control. A standard RS-232 cable with straight-through connections is required.
To this end, Ludlum Measurements, Inc. strives to supply the consumer of its goods with information regarding reuse and recycling of the many different types of materials used in its products.
Model 2200 Scaler Ratemeter Section Parts List Reference Description Part Number Model 2200 Scaler UNIT Completely Assembled Ratemeter Model 2200 Scaler Ratemeter 48-1651 BOARD Assembled Board 5167-413 Main Board, Drawing 167 x 413 MICRO 6.144 MHZ 01-5262 CRYSTAL C1-C3 100pf, 100V...
Model 2200 Scaler Ratemeter Appendix C – Scaler Printer Option Ludlum Measurements, Inc. sells a small 40-column serial-input impact printer for use with our new (year 2001) scaler or scaler/ratemeters. This small printer can print the results of each scaler count.
Model 2200 Scaler Ratemeter Appendix D – Drawings and Diagrams Main Board, Drawing 167 x 413 (4 sheets) Main Board Component Layout, Drawing 167 x 414 Amplifier Board, Drawing 167 x 392 Amplifier Board Component Layout, Drawing 167 x 393A...
Need help?
Do you have a question about the 2200 and is the answer not in the manual?
Questions and answers