ERROR INDICATORS AND TROUBLESHOOTING TIPS

ERROR INDICATORS
SYMBOL
Air plug disconnected.
Arm cuff not applied correctly.
Air is leaking from the arm cuff.
Cuff over inflated.
The arm cuff was inflated above 299
mmHg when inflating the cuff manually.
Movement during measurement.
Clothing is interfering with the
arm cuff.
Device error.
Sensor may be out of alignment.
Cuff is not applied correctly.
CAUSE
45
CORRECTION
Insert the plug securely. Refer to "Applying the
Arm Cuff".
Apply the arm cuff correctly. Refer to "Applying
the Arm Cuff".
Replace the cuff with the new one.
Repeat measurement. Remain still and do not talk
during measurement. Refer to "Taking a
Measurement".
If "E2" appears repeatedly, inflate the cuff manually
until it is 30 to 40 mmHg above your previous
measurement result. Refer to "Taking a Measurement".
Do not inflate the cuff above 299 mmHg. Refer to
"Taking a Measurement".
Repeat measurement. Remain still and do not talk
during measurement. Refer to "Taking a
Measurement".
Remove any clothing interfering with the arm cuff.
Refer to "Applying the Arm Cuff".
Contact Customer Service.
Contact Customer Service for repair information.
Refer to "Applying the Arm Cuff".

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Intellisense BP785 and is the answer not in the manual?

Questions and answers

george feth
March 23, 2025

My cuff started using three measurements with 60seconds in between, instead of one before giving me a reading. Did I accidently press wrong combination of buttons or something?

1 comments:
Mr. Anderson
March 23, 2025

The Omron Intellisense BP785 may be taking three measurements with 60 seconds in between instead of one because it calculates an average based on the last three measurements taken within 10 minutes. This helps improve accuracy by reducing the impact of temporary fluctuations in blood pressure due to factors like stress or movement.

This answer is automatically generated

Jeff Black
April 2, 2025

I purchased a omron bp 785n without Bloodpressure cuff or connector

User image 67ecc81a8c483
2 comments:
Mr. Anderson
April 2, 2025

Use only Omron authorized parts and accessories. Non-approved parts may damage the unit.

This answer is automatically generated