Replacing the probe base........................................8 Cleaning the probe base ........................................8 Cleaning the tonometer ........................................8 Replacing the batteries ......................................... 8 Returning the Icare tonometer for servicing /repair ................................9 Periodic Safety Checks .......................................... 9 Patents and copyrights ..........................................9 Symbols ..............................................9 Electromagnetic declaration ........................................
If servicing is necessary, contact qualified service personnel or your local Icare representative. WARNING Never spray, pour or spill liquid onto the Icare tonometer, its accessories, connectors, switches or openings in the chassis. Dry any liquid on the surface of the tonometer immediately.
English Icare tonometer TA01i manual CAUTION Federal law (U.S.) restricts this device to sale by or on the order of a physician. PARTS OF THE TONOMETER 1. Forehead support 2. Forehead support adjusting wheel 3. Display 4. Collar 5. Selector button 6.
English Icare tonometer TA01i manual probe to the patient’s cornea (see picture) should be 4-8 mm (1/6-1/3 inch). If necessary, adjust the distance by turning the forehead support adjusting wheel. Press the measurement button lightly to perform the measurement, taking care not to shake the tonometer.
English Icare tonometer TA01i manual MESSAGE STATE DESCRIPTION bAtt The batteries are low. Replace the batteries. E 01 The probe did not move at all. If this error message is repeated, turn the tonometer so that the collar faces down for a short time. If the error message is repeated, remove the probe base and replace it with new ones (see page 9).
The tonometer should only be opened by qualified service personnel. It contains no userserviceable parts, apart from the batteries and a probe base. The Icare tonometerrequires no routine servicing or calibration other than changing the batteries at least every 12 months or changing or cleaning the probe base.
Cleaning the tonometer WARNING Never spray, pour or spill liquid onto the Icare tonometer, its accessories, connectors, switches or openings in the chassis. Dry any liquid on the surface of the tonometer immediately. Icare TA01’s surfaces have been tested and found chemically resistant to the following liquids: •...
Icare TA01i tonometer. WARNING Use of any accessory or cable with the Icare TA01i tonometer other than those specified may result in increased emissions or decreased immunity of the Icare TA01i tonometer.
Page 10
Guidance and manufacturer’s declaration-Electromagnetic emissions TA01i is intended for use in the electromagnetic environment specified below. The user of the TA01i should assure that it is used in such an environment. RF emissions CISPR 11...
Page 11
Recommended separation distances between portable and mobile RF communications equipment and Icare TA01i Icare TA01i is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the IcareTA01i can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and Icare TA01i as recommended below, according to the maximum output power of the communications equipment.