Summary of Contents for National Instruments 2 SD Card Memory Module NI 9802
Page 1
OPERATING INSTRUCTIONS AND SPECIFICATIONS NI 9802 2 SD Card Memory Module...
Page 2
The National Instruments 9802 is an SD Card storage module that can read from and write to SD Cards. The NI 9802 is supported only in CompactRIO Note active chassis, such as the cRIO-910x. This document describes how to use the NI 9802 and includes specifications for the NI 9802.
Page 4
Substitution of components may impair Caution suitability for Class I, Division 2. For Zone 2 applications, install the system Caution in an enclosure rated to at least IP 54 as defined by IEC 60529 and EN 60529. Special Conditions for Hazardous Locations Use in Europe This equipment has been evaluated as Ex nA IIC T4 equipment under DEMKO Certificate No.
Page 6
descriptions of each of the LEDs on the NI 9802 and when it is safe to remove SD Cards from the NI 9802. 1 Present (Card 0) 2 Busy (Card 0) 3 Present (Card 1) Figure 1. NI 9802 Card Slots and LEDs NI 9802 Operating Instructions and Specifications 4 Busy (Card 1) 5 Card Slot Doors...
SD Card Slot Doors Do not open an SD Card slot door or remove an SD Card while the yellow LED for that card is lit or flashing. If you open an SD Card slot door while the yellow LED for that card is lit or flashing, the NI 9802 software acts to protect the file system of the SD Card.
Data Storage Characteristics Number of SD Cards ... 2 Storage ... 2 GB max storage for each Data transfer... 2 MB/s read/write for one MTBF ... 2,106,312 hours at 25 °C; Contact NI for Bellcore MTBF specifications Note at other temperatures or for MIL-HDBK-217F specifications.
Page 12
Europe (DEMKO)... Ex nA IIC T4 Environmental National Instruments C Series modules are intended for indoor use only but may be used outdoors if installed in a suitable enclosure. Refer to the installation instructions for the chassis you are using for more information about meeting these specifications.
Electromagnetic Compatibility This product is designed to meet the requirements of the following standards of EMC for electrical equipment for measurement, control, and laboratory use: • EN 61326 EMC requirements; Industrial Immunity • EN 55011 Emissions; Group 1, Class A •...
WEEE recycling centers and National Instruments WEEE initiatives, visit environment/weee.htm Where to Go for Support The National Instruments Web site is your complete resource for technical support. At everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers.
Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545 National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section on ni.com/legal for more information about National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies.
Need help?
Do you have a question about the 2 SD Card Memory Module NI 9802 and is the answer not in the manual?