Summary of Contents for Dr. Honle Dermalight 80 UV-B-311 nm
Page 1
PERATING NSTRUCTIONS ® dermalight psoracomb UV-B-311 nm – narrowband PO Box 154, Ballajura, Western Australia, Australia 6066 Tel: (+61) 08.9375.6447 Fax: (+61) 08.9375.5085 www.beatpsoriasis.com info@beatpsoriasis.com www.beatpsoriasis.com This document must be read carefully before using the dermalight80 product. Version 3.0 0123 0123...
Thank you for choosing the dermalight ® 80 from www.beatpsoriasis.com ® dermalight 80 is the original and most technically advanced phototherapy device available. It is both convenient and economical and been proven highly effective by various clinical trials for the treatment of many skin diseases. While several imitation products exist, rest assured that you are using the latest product available.
RICING OTICE FOR LIENTS Please note that all invoices and credit cards are charged in AUSTRALIAN dollars, or $AUD. Export orders are charged at the appropriate advertised AMERICAN dollar rates, converted to AUSTRALIAN using the daily international money exchange rate as published at www.xe.com Therefore, prices quoted in American are close...
AFETY NFORMATION ® Read the operating instructions carefully before using the dermalight 80 for the first time. All important safety information is marked with this symbol. Please follow these instructions carefully to prevent health risks and damage to property. dermalight ®...
Page 5
Additional safety instructions for doctors/clinics: The operator (doctor/clinic) has to check the careful operation of the personnel handling the equipment in regular intervals. ® Before commencing work, those people who are to work with the dermalight have to: • Observe the regulations governing safety and accident prevention. •...
ESCRIPTION 5.1. ENERAL dermalight ® 80 is a medical radiation appliance designed for mobile use. The emission spectrum of its lamps lies in the UVB narrow band range. ® Simple and universal use of the dermalight 80 is made possible by its low weight and handy shape.
PERATION ® Plug the dermalight 80 into a mains outlet (110 V, 60 Hz or 240v, 50Hz; please ® ensure appropriate power supply for your location). Operate the dermalight 80 only from properly installed socket outlets. With the rocker switch on the top of the ®...
Contra-indications: The following contra-indications are valid for all radiation units: • hyperthyroidism • lupus erythematosus • tumorous skin changes • diseased hyperphotosensitivity (e.g. xeroderma pigmentosa) • porphyrin disease • floride tuberculosis or other floride processes • accompanying medical treatment with potential skin damage or intensified photosensitivity •...
Page 9
Guide values for the maximum final exposure time (to complete the irradiation treatment) are: approx. 3-4 minutes for each irradiated area or approx. 2 times 20 minutes for the entire hair-covered scalp (f. ex. once in the morning and once in the evening). UV safety goggles guaranteeing 100% UV protection must be worn during treatment.
7.3. HERAPY RECOMMENDATION Consult your doctor for a recommendation on your initial treatment time. Example 1: Psoriasis of the hair-covered scalp Therapy: UV-B-311 nm therapy ® dermalight 80 with comb attachment Appliance: Patient's skin type: 2 (European average) Irradiation distance: 2- 3 cm (is kept automatically by the comb attachment) Duration of first exposure:...
Page 11
Example 3: Treatment of Vitiligo or Acne Distance: 3 cm, use the comb attachment as a distance holder. The comb attachment does not influence the treatment. Frequency: 3 times a week Maximum treatment time: 4 min 40 sec If skin irritations appear, decrease the treatment time of the following treatment by about 20 seconds.
AINTENANCE NFORMATION Daily: Check the filter glass for damage and dirt; check the housing for damage. Every 300 operation hours : Replace the lamps. Every 2 years: Maintenance by our Customer Service or a technician authorized by us (applies only to commercial users but is compulsory to comply with the German Medical Products Act).
8.2. LEANING THE Disconnect the appliance by unplugging the power-pack! Never wipe the hand lamp or the filter glass with a dry cloth. To clean dusty or dirty surfaces, wipe with a damp soft cloth or sponge (add a little dishwashing liquid to warm water).
ECHNICAL ® Type dermalight 80 UV-B-311 nm Order No. 19233 110v / 19158 240v Power supply #1, 110v 110 V ~ / 60 Hz (mainly USA) Power supply #2, 240v 240 V ~ / 50 Hz (European) Power consumption 20 VA Lamp 1 x UV-B-311 nm (PL-S 9W/01) Dimensions (L x W x H)
Page 16
Symbols: device of protection class II Attention, see operating instruction part of application: type B not ionised radiation drip proof IPX1 (Subject to technical alteration!) Therapy spectrum: Relative spectral radiation distribution (%) Wave length [nm]...
ARRANTY AND IABILITY In principle, the general terms of sale and delivery of Beat Psoriasis shall apply. These shall be submitted to the operator upon the signing of contract at the very latest. There shall be no warranty or liability claims in the event of personal injury or damage to property if these are attributable to one or more of the following causes: •...
Need help?
Do you have a question about the Dermalight 80 UV-B-311 nm and is the answer not in the manual?
Questions and answers