2. Calibration

2.1. General Calibration

Pressing and holding calibration switch press [POWER] key to go to calibration mode.
User can move to other mode by using [ZERO] key in the calibration mode.
User also moves to other sub-modes for each mode by using [TARE] key.
Please simply follow below procedure to move to other mode.
(1) Calibration Mode: Pressing and holding "Calibration Switch" press [POWER] key.
(2) It displays "CAL-0" after "CAL", and it blinks the version of scale three times.
(3) Selecting menu: press [TARE].
(4) ENTER(Setting) : [TARE] key
MODE
CAL 1
CAL 2
"UnLoad"
CAL 3
"MIDD"
"FULL"
"MIDD"
CAL 4
CAL 5
CAL 6
CAL 7
CAL 8
CAL 9
"Unit"
"CAPA"
CAL 10
"MCAPA"
"W-dP"
" 1 d "
"Dual"
CAL 11
Display normalized AD
Display Keypad infomation-
Weight Setting Mode
[TARE]
[TARE] after loading for 1/3 weight
[TARE] after loading for Full weight
[TARE] after loading for 1/3 weight
Option Setting ( Table 1 참조 )
Display filtered Raw AD
Function setting on each Key ( Table 2 참조 )
Battery calibration
Set calibration factor
[TARE]
select 0, 1 (0:kg, 1: lb)
[TARE]
select capacity
[TARE]
select mid-capacity
[TARE]
Select Decimal Point
[TARE]
Select division
[TARE]
Enable dual interval (0:disable, 1:enable)
Set nation(00 : OIML , 01 : NTEP , 02: KOREA)
9
Function
"END"
% Calibration
Gravity constant
[TARE]
[TARE]
[TARE]
[TARE]
[TARE]
Service Manual
ED
TARE
04/12/10