AVEA ventilator systems
Nasal CPAP / Nasal IMV Alarms
Note:
The Alarms Settings window is not available in this mode when the Rate is set to Off. Existing machine alarms and
safety systems will be maintained. During nCPAP support with Rate set to Off, certain alarms will be suspended.
Alarms suspended during nCPAP with Rate set to off
Time-based alarms
High Rate
I-Time Limit
I:E Limit
Apnea Interval
Low Ppeak
Range: 1 to 40 cmH
Default: 7 cmH
Note:
Not available in nasal CPAP mode unless Rate of one or greater is set.
Alarm threshold setting changes during nCPAP / IMV will not be retained for other ventilation modes following exiting
from nasal CPAP mode.
L2786 Rev. M
Chapter 7: Alarms and Indicators
Volume-based alarms
High Ve
High Vt
Low Vte
Low Ve
Volume Limit
O (Neonate setting – nCPAP / IMV)
2
O
2
Pressure alarms
High Ppeak
Ext High Peak Alarm
Low PEEP
Low Ppeak
Occlusion