Thank You! Thank you for your purchase of the SoClean 2 PAP disinfecting system. Please familiarize yourself with this manual and the operation of the unit. After a few simple steps, you’ll find that daily disinfecting of your PAP equipment has never been easier. The SoClean disinfecting process eliminates bacteria, mold, viruses and other pathogens, without the need for water or messy liquid chemicals.
Contraindications for Use: Do not use SoClean if a fragrant essence is used in the PAP water reservoir or if highly fragrant detergents are used to clean the PAP equipment. Discontinue use if acne or rash develops along mask contact line.
Included with your SoClean Removable hose slot plug SoClean filter AC power adapter ktech model cartridge KSAS0151200150D5 (installed) SoClean check valve with injection fitting and tubing (installed)
Important Supplies Check Valve (installed) Prevents water in your humidifier reservoir from backing up into your SoClean machine. If you can see water in the clear tube, replace the check valve assembly immediately. Replace every 6 months, or when the SoClean display shows the “Order Filter Kit”...
Understanding Your PAP Machine The SoClean is designed to connect to your PAP machine. This manual references some of the individual components of your PAP machine. Please use the diagram below to familiarize yourself with these terms before proceeding. Please note that there are many types of PAP machines. This diagram is for example reference only.
Use a fragrance-dish soap with 4 liters of water to wash your PAP equipment (mask, hose, and reservoir), then rinse thoroughly with clean water. Note: The SoClean uses no fluids in its operation. Pre-washing your equipment helps neutralize pre-existing odors. DO NOT immerse your SoClean in water.
Page 9
You may operate the SoClean with or without water in the humidifier reservoir. The SoClean will disinfect any water in the humidifier reservoir as long as the end of the injection tubing rests on or near the bottom of the reservoir, and below water level.
Page 10
Warning: Position your SoClean in a way that makes the power adapter accessible, so that you may unplug Plug your AC adapter into the back of your SoClean when necessary. the SoClean and then into a wall outlet. Proceed to set your timer.
Increase Status Indicator Light Your SoClean timer will allow you to completely automate your daily cleaning. The SoClean comes preset to start a 7 minute cleaning cycle at 10:00 am. First set the clock to match your current time. You can easily change these settings by following the steps on the next pages.
Your SoClean Timer & Display ... continued Setting Your Language Time Format Language Press the PlusandClockbutton. Set Current Time Set Current Time Press plusor minuto navigate to your Set Current Time Set Current Time language and then pressManual to select.
For those times when you wish to disinfect your equipment other than during your scheduled cycle, simply place your mask inside the SoClean like you normally would and press the button. The SoClean will then operate according to your preset run time.
Hose not present. Cycle will not run: present This message displays during the manual mode when there is no mask and hose in the SoClean, the lid is not properly closed, or the hose slot plug is missing or improperly seated. Order...
3. Insert your mask into the chamber allowing its attached hose to rest in the open hose slot. Note: The SoClean gives you the option of inserting your mask and hose from either side of the machine. To change sides,...
Page 16
SoClean will not operate if the hose slot plug is missing or improperly seated. Important: Even though the actual run time of the SoClean may only be 5 to 10 minutes, it is necessary to let the mask remain in the closed chamber for at least 2 more hours.
Care and Maintenance for your SoClean You may wipe down the SoClean with a damp cloth. The SoClean check valve and filter cartridge should be replaced every 6 months. A reminder message on the display will appear after approximately six months, depending on usage.
SoClean. If two hours have passed since the SoClean 2. Is the SoClean harmful to me or the completed a cycle, your PAP equipment is environment? No. The SoClean disinfects disinfected and ready for use.
Classification, Symbols, FCC Notice CLASSIFICATION Class lla According to 93/42/EEC MDD Annex IX,Sect lll,Rule 15 PAP Disinfecting device SYMBOLS FCC NOTICE Please note that changes or modifications not expressly Consult Manual approved by the party responsible for compliance could void the user’s authority to operate the equipment.