Dental - Miele G 7881 Operating Instructions Manual

Washer-disinfector
Table of Contents

Advertisement

Dental

A special holder AUF 1 (optional extra)
is required for processing hand pieces
and contra angles in top basket
O 177/1.
Separate operating instructions are
supplied with the holder.
Hand pieces and contra angles with
fibre optic cables should be monitored
frequently as fibre optic cables can
deteriorate more quickly.
Care should be taken immediately after
treating a patient that as little dental
cement or amalgam remains on the
instruments as possible. If necessary
wipe off excess with a swab.
Manual cleaning can be carried out if
required after thermal disinfection in
accordance with national health and
safety regulations.
Rhodium plated specula must be
arranged in such a way that they
cannot suffer mechanical damage.
Not all specula with glass are
suitable for machine processing.
Areas of application
Lower basket /carrier U 874/2
for inserts, e.g. E 130 and E 379.
Lower basket /carrier U 874/2
for inserts, e.g. E 131/1 for 5 mesh
trays with support E 146 or 5 mesh
trays with support E 363 for double
ended instruments.