Installation Conditions - Hitachi TM3000 Instruction Manual

Tabletop microscope
Table of Contents

Advertisement

1.4 Installation Conditions

Install the system in a place that meets the following conditions.
Item
General
1. Level location
requirements
2. Away from water spill
3. Away from external shock
4. Away from heat sources
5. Away from a strong magnetic field (vertically from system: 15 nT,
6. Absence of a strong electric field
7. Absence of dust particles
8. Absence of vibrations
Vibration
Frequency ≦ 7 Hz, 4.0 µm p-p or less
10 Hz < frequency, 2.0 cm/s
Magnetic field
Noise
75 dB or less (frequency ≦ 150 Hz) [C characteristics: 72 dB max.
(frequency ≦ 150 Hz)]
Ambient
15 to 30°C (
temperature
Ambient
45 to 70[%RH]. (No condensation)
humidity
Main unit
Single-phase AC 100 to 240 V (Min. 90 V, Max. 250 V), 500 VA,
power supply
50/60 Hz
3P outlet (grounding resistance of less than 100 Ω)
Installation
Recommended: 700 mm (H) x 1200 mm (W) x 800 mm (D) or larger
area
Minimum: 700 mm (H) x 1000 mm (W) x 800 mm (D) or larger
Table 1.4 Installation conditions
horizontally from system: 5 nT)
7 Hz < frequency ≦ 10 Hz, 1.5 µm p-p or less
Allowable stray magnetic field for the TM3000
Accelerating voltage: 5 kV
Magnification: 2 kx
D:4.5 mm
Accelerating voltage: 15 kV
Magnification: 5 kx
D:4.5 mm
t=±2.5°C/h or less)
Installation conditions
2
or less
AC magnetic field
(nT)
Horizontal
Vertical Horizontal
151
243
116
148
1 - 4
DC magnetic field
fluctuation (nT)
(Peak-to-Peak)
Vertical
1065
1717
818
1045