Before installation, please read all instructions. The DewPro
is designed to be mounted to pressurized systems. Take all
necessary precautions when mounting or
removing the DewPro.
Mount the DewPro vertically whenever possible to prevent
particles or condensation from entering the bypass.
Mount the DewPro after a shut-off valve to depressurize the
DewPro when removing it from the process pipe in case of
maintenance or field compensation.
Do not over-tighten! The outlet fitting is connected to the
bypass block with a G 1/4 straight thread (with gasket)
which will seal if the fitting is simply hand-tightened. When
connecting an external device, counter the fitting with a
second wrench when tightening. If the inlet is equipped
with a G 1/2 straight thread and gasket, the seal is
obtained by simply hand-tightening the DewPro.
If you are installing the DewPro into a pressurized system
(up to 10 bar), depressurize the system before installing or
removing the sensor. Pressurized systems require a
stainless steel compression fitting.
Caution!
Caution!
Caution!
November 2005
2-1