Model 654A Test Oscillator - HP 654A Operating And Service Manual

Test oscillator
Hide thumbs Also See for 654A:
Table of Contents

Advertisement

_
Section I
Model 654A
Figure 1-1. Model 654A Test Oscillator
lable I- I. Specifications
Frequency range: 10 Hz to 10 MHz in 6 bands.
Frequency accuracy:
100 Hz (on X100 RANGE) to 5 MHz: +/-2%
10 Hz to 100 Hz: +/-3%
5 MHz to 10 MHz: +/-4%
Level flatness(+10 dBm and 0 dBm): +/-0.5% referenced
to level at 1 kHz from 10 Hz to 10 MHz for un-
balanced outputs, 10 Hz to 5 MHz for 135 ohm and
150 ohm outputs, and 10 Hz to 1 MHz for 600 ohm
output.
Attenuator
Range: 99 dB in 10 dB and 1 dB steps.
Accuracy: +/- 1.5% (0.15 dB) except +/- 10% ( 1 dB) at
output levels below 60 dBm at frequencies greater
than 300 kHz.
Amplitude control: greater than 2 dB.
Amplitude accuracy: +/- 1% for 90 days (at
1 kHz,
+10 dBm level with meter centered).
Meter tracking: +/-0.05 dB.
Balance (on balanced impedances) when measured by the
procedure given in Paragraph 5-28: greater than 50 dB
for frequencies from 10 Hz to 1 MHz, greater than
40 dB to 5 MHz.
Distortion (THD)
10 Hz to 1 MHz: greater than 40 dB below funda-
mental.
1 MHz to 10 MHz: greater than 34 dB below funda-
mental.
Hum and noise: greater than 70 dB down at full output.
Table 1-2. General Information
Output
impedance: 50 ohm unbalanced, 75 ohm un-
balanced, 135 ohm balanced, 150 ohm balanced and
600 ohm balanced.
Output level: + 11 dBm to -90 dBm, 10 dB and 1 dB steps
with adjustable +/- 1 dB meter range; calibrated for
each impedance.
Meter resolution: 0.02 dB.
Output connectors: BNC. Maximum voltage which can be
applied to the output: less than +/-3 V peak.
Counter output: greater than 0.1 V nus into 50 ohm,
BNC connector.
Operating temperature: 0°C to +55°C (32°F to 130°F).
Power: 115 V or 230 V +/- 10%, 48 Hz to 440 Hz, 30 W
nominal, 35 W max.
1-2

Advertisement

Table of Contents
loading

Table of Contents