Main Assembly Theory
Family Acquisition Differences
The following explains the differences in the three
models that comprise the rest of this family, the HP 54540A, 54522A, and 54520A.
e The HP
54540A oscilloscope has four channels like the HP 54542A except that its sample
rate capability is 500 MSa/s with four channels running and 1 GSa/s with two channels
running. The acquisition architecture is identical to the HP 54542A.
The HP 54522A has the sample rate capability of the HP 54542A, 2 GSa/s, but only two
signal channels. The third channel position is unused and the fourth channel position is
used for an External Trigger. The external trigger uses an attenuator/preamp for signal
conditioning and has some of the ranging capability of a channel but no acquisition circuitry
(there is no ADC hybrid). The output of the external trigger attenuator/preamp is fed to
the analog trigger just like a channel.
The HP 54520A has the two-channel configuration of the HP 54522A with 500 MSa/s
sample rate capability on both channels.
8-9