The Frame menu includes frame-specific service
menus.

2.1 Memory

A service menu to check the status of the
memory used in the CPU board of the monitor.
Test Memory tests the condition of the
EEPROM/Flash memory component of the CPU
board. If the result of the test is Fail, see section
"Error messages" in chapter "CPU."
Test SRAM tests the Static RAM memory of the
CPU board in a similar way as the
EEPROM/Flash memory. If the result of the test
is Fail, see section "Error messages" in chapter
"CPU."
Real-time clock test is run at every start up and
also during the operation of the monitor. If the
result of the test is Fail, the battery for the SRAM
timekeeper should be replaced.