6. Maintenance and Safety Inspection

1) After using the probe, the Magnetic sensor attachment and
the Spacer for EZU-RV2S, it should be cleaned and disinfected
and Sterilization following "5. Cleaning, High Level
Disinfection and Sterilization of the Probe, the Magnetic
sensor attachment and the Spacer for EZU-RV2S", then store
it in a cool and dark place avoid high temperature and
humidity, direct sunlight.
2) Visually inspect the surface of the probe head, housing, cable and
connector for any crack, scratch or denaturalization. If you find
any damage, do not use the probe and contact a service support
immediately.
3) Visually inspect the surface of the Magnetic sensor attachment and
the
Spacer
denaturalization. If you find any damage, do not use them and contact
a service support immediately.

7. Safety Precautions

1) Never use the probe if the probe head, housing or cable are cracked
or damaged.
2) Warning is case of using probe covers which latex is contained to.
The latex may cause such allergic reactions as itching, rubor,
urticaria, swelling, fever, anhelation, wheezing, and depression
of blood pressure, shock and so on. For the patients suspected of
latex allergy, do not use the latex-containing medical devices.
If you observe any of above mentioned symptoms in your patient during
the operation, stop the use of the latex-containing medical devices
immediately and take an appropriate treatment to the patient.
3) In case of using sterile probe covers, probe covers are packaged
sterile and are single-use only. Do not use if integrity of packaging
is violated or if expiration date has passed.
1) The probe connector is not waterproof. Do not allow liquid to contact
the connector.
2) Do not drop, hit or bent the probe.
3) Use only water, detergents and disinfectants in the suppliers list.
Between uses store the probe holder of scanner.
4) Under sterile condition use appropriate protection for probe and
cable. Some Latex material may create allergic reactions.
5) The probe and the biopsy attachment are not delivered disinfected
or sterilized.
for
EZU-RV2S
for
any
crack,
WARNING
CAUTION
-32-
deformation
or
Q1E-EP1245