Back End Processor Checks; Operator Panel Test; Peripheral Checks - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

4-3-30 Back End Processor checks

4-3-31 Operator Panel Test

4-3-32 Peripheral checks

4-3-32-1 Printer checks
4-3-33 Turn OFF Power to Vivid T9/Vivid T8
4-3-34 Mechanical Functions Checks
4-3-34-1 Operator Panel Movement
Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
If all the previous tests have been passed successfully, the
Back End Processor is most likely OK.
If the system seems to be operating erratically, please refer
to
'Diagnostics/Troubleshooting' on page
The Operator Panel is tested when the is powered up as
part of the start-up scripts, run at every start-up.
For more info, please refer to
on page
7-1.
The internal printer is controlled from the P1 and P2 keys on the
Vivid T9/Vivid T8 's Operator Panel.
The factory default is:
P1 for the UP-D711MD printer
P2 for the whole screen secondary capture to clipboard
See 4-2-3 'Power off' on page 4-8 for more information.
Please refer to:
4-2-4 'LCD Monitor position adjustment' on page 4-9
Functional checks
7-1.
'Diagnostics/Troubleshooting'
4-35