Cleaning And Replenishing The Cuvette Wash Bottle - Siemens ADVIA 2400 Maintenance Procedures

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

Figure 5-17. Isotonic Saline diluent bottle
1. Lift the cover from the saline diluent bottle (
CAUTION
Note the bottle position on the shelf, to avoid mixing up the fluid bottles.
2. Empty the remaining contents of the bottle.
3. Rinse the bottle with deionized water and drain well.
4. Refill the bottle with 0.9% saline diluent.
5. Replace the bottle in the same position on the shelf in the cabinet.
6. Replace the cover of the diluent bottle.
NOTE
Make sure that the Teflon tube and filter holder are located at the bottom of the
dilution bottle.
7. Prime the fluid lines:
NOTE
If you are cleaning the detergent or cell conditioner bottles at this time, you can
prime all the fluid lines at once.
a. At the Operation Panel, select
b. At the PRIME Settings dialog box, select
c. Select

Cleaning and replenishing the cuvette wash bottle

Materials required:
• Deionized water
• Cuvette detergent (wash solution)
Time: 10 minutes
Analyzer mode: READY
Maintenance
of the number of times fields.
Execute
.
1 ISE buffer bottle
2 RRV (Reaction) bath oil
3 Isotonic saline diluent bottle
4 Cuvette detergent bottle
5 Cell conditioner bottle
3
), and remove the bottle.
.
Prime
Prime 2
, then type
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
10