9 Reference information
9.1 Specifications
Environmental conditions
Parameter
Operating conditions
Transportation and stor-
age temperature
Lighting
Where to use
Max. altitude, operation
Overvoltage category
Pollution degree of the in-
tended environment
Degrees of protection
provided by enclosure
Free space required
around
Amersham Typhoon
scanner
Placement
134
Data
Temperature
18°C to 28°C
Humidity
20% to 70%, RH (no dew condensa-
tion)
Temperature
-20°C to 70°C
Humidity
5% to 95% (no dew condensation)
It is recommended to lower the lighting level to about
20 lux when moving a sample from the cassette into
Amersham Typhoon after exposure.
Indoor use only, out of direct sunlight or brightly lit sur-
roundings.
2000 m above sea level
Transient overvoltage category II
Pollution Degree 2
IP21
Front of instrument (the operation panel side): 70 cm
Right
10 cm
Left
10 cm
Rear
10 cm
Top
10 cm
A stable laboratory bench with a load capacity of 2940
N (300 kg), or higher.
The bench must be level within two degrees.
Amersham Typhoon Operating Instructions 29193226 AC