To operate the reagent gas flow control module

For a video demonstration of the gas flow control module, see the
HP 5973 MSD Reference Collection CD-ROM.
Flow control module state diagram:
Result
Gas A flows
Control panel lights (LEDs)
Gas A (green)
On
Gas B (amber) Off
Purge (red)
Off
Gas Off (red)
Off
Valve state
Valve A
Open
Valve B
Closed
On → setpoint
MFC
Isolation valve Open
Gas B flows
Purge
with Gas A
Off
On
On
Off
Off
On
Off
Off
Closed
Open
Open
Closed
On → setpoint
On → 100%
Open
Open
7R RSHUDWH WKH UHDJHQW JDV IORZ FRQWURO PRGXOH
Purge
Pump out
with Gas B
flow module
Off
Off
On
Off
On
On
Off
On
Closed
Closed
Open
Closed
On → 100%
On → 100%
Open
Open
 CI operation
Standby, vented,
or EI mode
Off
Off
Off
On
Closed
Closed
Off (→0%)
Closed
61