Application technology
Preparing the
load
Wet loading
38
– Only reprocess small items and micro components in special
inserts, mesh trays with lids or mesh inserts.
– Plastic items must be thermally stable.
For validation purposes it is essential to follow the loading
instructions given on the template.
Observe the further information given in the following sections as
necessary depending on the area of application.
Empty the load before sorting
Ensure that no acid or solvent residues, especially hydrochloric
acid or chlorides, get inside the wash cabinet.
Dismantle the items for washing where possible according to the
manufacturer's instructions.
Follow the instructions of the instrument manufacturer regarding
pre-cleaning and pre-treatment as necessary.
Thoroughly rinse items which have been pre-treated with
chemicals.
Chemically pre-treated items must be rinsed thoroughly by hand or
using the Rinse program before reprocessing in the washer-
disinfector to avoid a significant build-up of foam.