Telemetry Functions at the Information Center
To silence the
sound
Viewing
Device
Location in
the Patient
Window
(optional)
Viewing
Device
Location
History
(optional)
Patient
Configurable
Settings in
Telemetry
Setup
9-4
Telemetry Functions at the Information Center & TeleMon
Step
Action
1
Press and hold the Check button for 6 seconds to turn off the sound.
OR
Remove the batteries.
You can see the location of a transceiver in the Patient Window. The Device
Location information is identified in the Patient Window by a compass icon
followed by the location name configured by your hospital for your system. If
the location of the transceiver changes, the Patient Window is updated within 5
seconds of the location change.
You can view the location history for a particular transceiver in the Device
Location History field in the Telemetry Setup window. The field displays the
five most recent Device Location descriptions in ascending order and updates
every time there is a change in location for the device. The total timespan of the
log is 60 minutes.
If there is a change in location while viewing the Telemetry Setup
Note—
window, you must re-enter Telemetry Setup to see the change, as it does not
update automatically.
The Telemetry Setup window enables you to configure the transceiver for
patient-specific settings. All patient-specific settings will be reset to the unit
defaults upon patient discharge. To access the window, from the Patient
Window click Telemetry Setup.
The following settings can be adjusted in this window.