Chapter 4: Installation
Figure 8: Wiring Option — Binary Input using Analyzer's Internal Supply
Figure 9: Wiring Option — Binary Input using Analyzer's External Supply

Installing the Analysis Complete Output Connection

The analysis complete (binary output) on TB3 may be used as a control signal to a sample sequencer device.
Once the binary output is wired properly, the Analyzer will automatically provide a one-second contact closure
at the end of every sample analysis.
No menu configuration is necessary. The binary output is active in the Online, Check Standard, and Grab modes.