Limit Function Setup (Cont.) - Hitachi AJ-CE series Operation Manual

High-precision tuning fork electronic balance
Table of Contents

Advertisement

8.1 Limit Function Setup (cont.)
2
Selecting a function item
Press the | Function | key.
The display changes to the next item
[Limit Function].
3
Setting the limit function
?.S£L
•'
Press the | Zero/Tare | key to set the
value on the rightmost side to [1].
Limit function
operations
4
Setting the judgement condition
l.Co. I
Press the | Function I key.
The display changes to [Judgement
Condition].
Press the | Zero/Tare"] key to select
the desired condition.
Judged at all times
Judged when stable
5
Setting the judgement range
aa.L
aa.L
u
Judge entire range
Press the | Function | key.
The display changes to [Judgement
Range].
Press the I Zero/Tare~l key to select
the desired choice.
Judge beyond +5
6
Setting the number of judgement points
Press the [ Function"] key.
The display changes to [Judgement
Points].
Press the | Zero/Tare~] key to select
the desired choice.
One point set
Upper/lower limits set
7
Terminating the function setup
IDOOt
Press the | Set I key.
The balance terminates the function
setup and returns you to measurement
mode.
23

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ajh-ce series

Table of Contents