Electrical Troubleshooting - Honeywell F57A Manual

Flush-mount commercial electronic air cleaner
Hide thumbs Also See for F57A:
Table of Contents

Advertisement

F57A,B FLUSH-MOUNTED COMMERCIAL ELECTRONIC AIR CLEANER
3. Check the cell for short circuits using an ohmmeter. See
Fig. 16. Check the resistance between the frame of the
cell and both the ionizer and the collector contacts. In
each case, the resistance should be infinite.
REPLACING AN IONIZER WIRE.
SPRING
CONNECTORS
PRESS
DOWN
IONIZER
WIRE
TWO EYELETS HOLD IONIZER WIRE TO CELL.
1
Fig. 15. Install new ionizer wire by hooking
eyelets over spring connectors.
COLLECTOR
TERMINAL
Fig. 16. Use an ohmmeter to check
the cells for short circuits.
IONIZER
EYELETS
WIRE
1
NEEDLENOSE
PLIERS
M1540B
IONIZER
TERMINAL
COLLECTOR
TERMINAL
M6155
Modification to Reduce Ozone
Odor
The electronic air cleaner generates a trace amount of
ozone in normal operation. During the first week or two of
operation, the amount may be higher because of sharp
edges on some of the new high voltage metal parts.
However, normal use and the first wash quickly dull the
edges.
The average person can detect the odor of ozone indoors in
concentrations as low as 0.003 to 0.010 parts per million
(ppm). The electronic air cleaner contributes 0.005 to 0.010
ppm of ozone to the indoor air. The U.S. Food and Drug
Administration and Health and Welfare Canada recommend
that indoor ozone concentration should not exceed 0.050
ppm. As a comparison, the outdoor ozone level in major
cities ranges from 0.020 ppm to 0.040 ppm and even higher.
However, if desired, the ozone generated by the air cleaner
can be reduced in one of two ways:
1. Install the optional activated carbon filter.
2. Open the grille and move the electronic air cleaner
performance switch to the low setting. Close and
relatch the grille. The cleaning efficiency is now
reduced 5 to 15 percent, depending on airflow.
NOTE: Be sure the switch is set to either high or low
position, not between or the unit may not run.
Be careful not to reset the switch when closing
the cover.
ELECTRICAL
TROUBLESHOOTING
WARNING
Electrical Shock Hazard
Can cause personal injury or equipment damage.
The following procedures expose hazardous live
parts. Disconnect power supply between checks and
proceed carefully.
CAUTION
The following instructions are for use by only qualified
personnel.
Tools and Equipment
Troubleshooting the electronic air cleaner requires:
• Needlenose pliers for stringing ionizer wires and inserting
edge connectors.
• Test meter with 15 kV dc probe.
Troubleshooting Procedure
The electronic air cleaner Troubleshooting chart, Fig. 17
shows how to quickly isolate a problem in the air cleaner.
See Figs. 18 & 19 for electrical schematics.
15
68-0083-7 Revised 11-07

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

F57b

Table of Contents