CAUTION
The wired foot switch must always remain connected to your X-ray system and be available for clinical
use when using the wireless foot switch. If image acquisition cannot be started using the wireless foot
switch, you can continue the procedure with the wired foot switch.
NOTE
Ensure that the wireless foot switch battery is sufficiently charged before using the foot switch. For
information about battery charge level, see
depleted during a procedure, the foot switch will stop working. In this case, use the wired foot switch.
Alternatively, connect the charger to the foot switch and continue to use it. Take care not to damage
the cable of the charger when moving equipment around the examination room (for example, when
moving carts or beds).
NOTE
If you disconnect the charger while X-ray is being performed (while a pedal is being pressed), image
acquisition stops and should be restarted, if needed, by releasing and pressing the pedal again.