C. OUTPUT CIRCUIT AND DC OFFSET
C1.
Preliminary settings of controls:
PM 5775-( PM 5776)
REPETITION TIME, S1
10 us
FINE, RV1
12 o'clock
DELAY, S82
100 ns
FINE, RV2
fully counter-clockwise
WIDTH, 83
1 us
FINE, RV3
fully counter-clockwise
CONT. GATED, S6
CONT
SINGLE DOUBLE, $7
SINGLE
AMPLITUDE IN 50 Q, S4
—3V
Vernier, RVS
MAX.
DC OFFSET IN 50 Q, RV4
OV
NORM. INV., S8
INV.
PM 3400
mV/cm
50, CAL
TIME/cm
2 ps
Connect socket PULSE OUT 50 Q, P3, to one of the oscilloscope inputs via a 50 Q
coaxial cable and a 20 dB attenuator (e.g. General Radio GR874 G20). Connect
socket SYNC. OUT, P2, to the TRIGG. input of the oscilloscope via a 50 Q
coaxial cable and a 20 dB attenuator (e.g. Texscan type FP 50, BNC Outline A).
Bias Z330
— Observe a flat pulse top on the oscilloscope screen. If required, adjust with
RV301. Fig. X-4 shows unacceptable deviation.
correct shape
set with
RV302
a
Q.=ta=
unacceptable
deviation
-3V-
MAE&499
Fig. X-4. Pulse diagram