Page 2
Operating Instructions Landcal Blackbody Source Type P80P SAFETY INFORMATION This product complies with current European directives relating to electromagnetic compatibility and safety (EMC directive 89/336/EEC; Low voltage directive 73/23/EEC). EN 61010-1 Symbol identification Symbol Publication Description IEC 417, Nº 5031 Direct current IEC 417, Nº...
Operating Instructions Landcal Blackbody Source Type P80P Contents INTRODUCTION DESCRIPTION SPECIFICATION ELECTRICAL SUPPLY COMMISSIONING Inspection on receipt Connections to mains Supply Heating up/cooling down of the source Using the RS232 serial interface port USING THE SOURCE Introduction Measuring sensor (Platinum resistance thermometer) - if supplied...
To make the lining up of LAND fixed installation radiation thermometers simpler, an optical bench assembly is offered as an optional extra. When the source is stood on the optional transportation case, the bench to mid target dimension equals that of the optical bench assembly.
Type P80P DESCRIPTION The LANDCAL blackbody source type P80P comprises a cylindrical closed end tube (cavity) approximately 160mm/6.3in long with an internal diameter of 50mm (2.0in). The cavity is manufactured from Aluminium which is blackened and the closed end is angled at 120° to increase the emissivity value.
Operating Instructions Landcal Blackbody Source Type P80P SPECIFICATION Maximum working temperature: 80°C (175°F) Recommended temperature range: -10 to 80 °C (15 to 175°F) Stability: With the source controlling at temperature the radiance temperature will vary by less than ±0.5K (±1°F) over a 30 minute period.
Landcal Blackbody Source Type P80P ELECTRICAL SUPPLY The P80P is supplied with a removable electrical supply cable. The cable has an IEC style connector on one end that mates to an interated power entry module. The colour code for the cable is:...
Operation of the source at below ambient temperature The LANDCAL Blackbody Source type P80P uses Peltier modules to cool the source to temperatures as low as -10°C/ 15°F. However the modules are rated as only achieving a 30°C (55°F) below ambient cooling effect. This means that the final cooling temperature achieved and the length of time it takes to achieve this temperature will be dependent on the ambient temperature.
±2K. To make the lining up of LAND fixed installation radiation thermometers simpler an optical bench assembly is offered as an optional extra. When the source is stood on the optional transportation case, the bench to mid target dimension equals that of the optical bench assembly.
If difficulty is experienced in writing procedures, LAND Infrared would be pleased to offer guidance as to what calibration conditions should be adopted for LAND products.
P80P source and/or the standard radiation thermometer should be returned to LAND every 1 to 3 years for recertification. UKAS Certificates of Calibration are available from Land Infrared, U.K. Certificates of calibration traceable to N.I.S.T. are available from Land Infrared, U.S.A.
Operating Instructions Landcal Blackbody Source Type P80P EUROTHERM TEMPERATURE CONTROLLER User Guide Output light Measured temperature EUROTHERM Setpoint temperature Page Scroll Down Fig. 3 Eurotherm Controller - front panel controls CA970253/a When switched on, the controller lights up, goes through a short test routine, and then displays the measured temperature and starts to control.
Operating Instructions Landcal Blackbody Source Type P80P Altering the Setpoint Press either the key once to display the setpoint. Down (ii) Use the key to adjust the setpoint value. Down The display returns to the measured temperature when no key is pressed for 0.5 seconds.
Operating Instructions Landcal Blackbody Source Type P80P Altering the communication address Press the page key until cmS is displayed. (ii) Press the scroll key until Addr (address) is displayed. (iii) To alter the value press the up or down key.